Your browser doesn't support javascript.
loading
Electrospray ionization from nanopipette emitters with tip diameters of less than 100 nm.
Yuill, Elizabeth M; Sa, Niya; Ray, Steven J; Hieftje, Gary M; Baker, Lane A.
Afiliación
  • Yuill EM; Department of Chemistry, Indiana University , 800 E. Kirkwood Avenue, Bloomington, Indiana 47405, United States.
Anal Chem ; 85(18): 8498-502, 2013 Sep 17.
Article en En | MEDLINE | ID: mdl-23968307
ABSTRACT
Work presented here demonstrates application of nanopipettes pulled to orifice diameters of less than 100 nm as electrospray ionization emitters for mass spectrometry. Mass spectrometric analysis of a series of peptides and proteins electrosprayed from pulled-quartz capillary nanopipette emitters with internal diameters ranging from 37 to 70 nm is detailed. Overall, the use of nanopipette emitters causes a shift toward the production of ions of higher charge states and leads to a reduction in width of charge-state distribution as compared to typical nanospray conditions. Further, nanopipettes show improved S/N and the same signal precision as typical nanospray, despite the much smaller dimensions. As characterized by SEM images acquired before and after spray, nanopipettes are shown to be robust under conditions employed. Analytical calculations and numerical simulations are used to calculate the electric field at the emitter tip, which can be significant for the small diameter tips used.

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: Anal Chem Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: Anal Chem Año: 2013 Tipo del documento: Article País de afiliación: Estados Unidos