Evaluation of Beckman Coulter DxI 800 immunoassay system using clinically oriented performance goals.
Akbas, Neval; Schryver, Patricia G; Algeciras-Schimnich, Alicia; Baumann, Nikola A; Block, Darci R; Budd, Jeffrey R; Gaston, S J Stephen; Klee, George G.
Afiliación
  • Akbas N; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
  • Schryver PG; Department of Health Sciences Research, Mayo Clinic, Rochester, MN, USA.
  • Algeciras-Schimnich A; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
  • Baumann NA; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
  • Block DR; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
  • Budd JR; Beckman Coulter, Chaska, MN, USA.
  • Gaston SJ; Beckman Coulter, Chaska, MN, USA.
  • Klee GG; Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA. Electronic address: klee.george@mayo.edu.
Clin Biochem ; 47(16-17): 158-63, 2014 Nov.
Article en En | MEDLINE | ID: mdl-25130957
ABSTRACT

OBJECTIVES:

We evaluated the analytical performance of 24 immunoassays using the Beckman Coulter DxI 800 immunoassay systems at Mayo Clinic, Rochester, MN for trueness, precision, detection limits, linearity, and consistency (across instruments and reagent lots).

METHODS:

Clinically oriented performance goals were defined using the following

methods:

trueness-published desirable accuracy limits, precision-published desirable biologic variation; detection limits - 0.1 percentile of patient test values, linearity - 50% of total error, and consistency-percentage test values crossing key decision points. Local data were collected for precision, linearity, and consistency. Data were provided by Beckman Coulter, Inc. for trueness and detection limits.

RESULTS:

All evaluated assays except total thyroxine were within the proposed goals for trueness. Most of the assays met the proposed goals for precision (86% of intra-assay results and 75% of inter-assay results). Five assays had more than 15% of the test results below the minimum detection limits. Carcinoembryonic antigen, total thyroxine and free triiodothyronine exceeded the proposed goals of ±6.3%, ±5% and ±5.7% for dilution linearity. All evaluated assays were within the proposed goals for instrument consistency. Lot-to-lot consistency results for cortisol, ferritin and total thyroxine exceeded the proposed goals of 3.3%, 11.4% and 7% at one medical decision level, while vitamin B12 exceeded the proposed goals of 5.2% and 3.8% at two decision levels.

CONCLUSIONS:

The Beckman Coulter DxI 800 immunoassay system meets most of these proposed goals, even though these clinically focused performance goals represent relatively stringent limits.