Rapid exchange ultra-thin microcatheter using fibre-optic sensing technology for measurement of intracoronary fractional flow reserve.
EuroIntervention ; 11(4): 428-32, 2015 Aug.
Article en En | MEDLINE | ID: mdl-26013582
ABSTRACT

AIMS:

The present report describes a novel coronary fractional flow reserve (FFR) system which allows FFR assessment using a rapid exchange microcatheter (RXi). METHODS AND

RESULTS:

The RXi microcatheter is compatible with standard 0.014" coronary guidewires facilitating lesion negotiation and FFR assessment in a wide range of coronary anatomies. In case of serial lesions, a microcatheter would have the important advantage of allowing multiple pullbacks while maintaining wire access to the vessel. The RXi is a fibre-optic sensor technology-based device. This technology might allow reduction in signal drift. The RXi microcatheter's fibre-optic sensor is located 5 mm from the distal tip. The microcatheter profile at the sensor site is 0.027"0.036". The segment of the catheter which is intended to reside within the target lesion is proximal to the sensor and has dimensions decreased to 0.020"0.025"; these dimensions are comparable to a 0.022" circular-shaped wire.

CONCLUSIONS:

The RXi microcatheter FFR system represents a novel technology that could allow easier lesion negotiation, maintaining guidewire position, facilitating pullbacks for assessment of serial lesions and simplifying the obtainment of post-intervention FFR measurements. The optical sensing technology could additionally result in less signal drift. Further investigations are required to evaluate the clinical value of this technology fully.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Cateterismo Cardíaco / Reserva del Flujo Fraccional Miocárdico / Tecnología de Fibra Óptica / Catéteres Cardíacos / Microcirculación Tipo de estudio: Evaluation_studies / Prognostic_studies Límite: Humans Idioma: En Revista: EuroIntervention Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Año: 2015 Tipo del documento: Article País de afiliación: Países Bajos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedad de la Arteria Coronaria / Cateterismo Cardíaco / Reserva del Flujo Fraccional Miocárdico / Tecnología de Fibra Óptica / Catéteres Cardíacos / Microcirculación Tipo de estudio: Evaluation_studies / Prognostic_studies Límite: Humans Idioma: En Revista: EuroIntervention Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Año: 2015 Tipo del documento: Article País de afiliación: Países Bajos