  • Patteet L; *Toxicological Centre, University of Antwerp; †Laboratory for TDM and Toxicology, ZNA Stuivenberg; ‡Collaborative Antwerp Psychiatric Research Institute (CAPRI), Faculty of Medicine, University of Antwerp; §Psychiatric Hospital Broeders Alexianen, Boechout; and ¶Psychiatric Centre Sint-Norbertushuis, Duffel, Belgium.
Ther Drug Monit ; 38(1): 87-97, 2016 Feb.
Article en En | MEDLINE | ID: mdl-26349083
ABSTRACT

BACKGROUND:

Oral fluid (OF) is an interesting alternative for conventional blood testing in therapeutic drug monitoring. OF can be used for screening but its value for quantification has to be established.

METHODS:

To evaluate the value of OF for quantification of 11 commonly used antipsychotics (APs) and 5 metabolites, an ultra-high performance liquid chromatography-tandem mass spectrometric method was validated. OF was obtained from psychiatric patients using a Quantisal collection device. OF to serum concentration ratios were determined, taking into account the exact volume of collected OF.

RESULTS:

Linearity was evaluated at 7 or 8 calibration levels. Accuracy criteria were fulfilled, except for pipamperone (PIP) at quality control (QC) low. The intraday precision ranged 0.88%-14.73% and interday precision ranged 1.92%-16.17%. The mean recovery from the collection pad was 37.1% at QC low and 40.3% at QC high for 1 mL of collected OF; for 0.5 mL collected OF mean recovery was 35.0% at QC low and 37.3% at QC high. When 0.1 mL OF was collected, recovery data were unreliable. Mean absolute matrix effect was 101.1% (82.0%-120.0%). OF patient samples (n = 89) containing 269 APs and metabolites were acquired and the mean volume of collected OF was 0.562 mL (0.057-1.232 mL). The OF to serum ratios were above 1 for all APs (1.54-28.50), except for aripiprazole (0.21) and zuclopenthixol (0.66). A broad range of calculated ratios for all APs was obtained.

CONCLUSIONS:

This validated ultra-high performance liquid chromatography-tandem mass spectrometric method can be used to reliably quantify APs in OF, even when recovery is low. Because the correlation between OF and serum concentrations was low and in addition results were highly variable, it can only be concluded that OF is a potentially interesting matrix, particularly for screening for noncompliance.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Antipsicóticos / Cromatografía Líquida de Alta Presión / Monitoreo de Drogas / Espectrometría de Masas en Tándem Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Ther Drug Monit Año: 2016 Tipo del documento: Article País de afiliación: Bélgica

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Antipsicóticos / Cromatografía Líquida de Alta Presión / Monitoreo de Drogas / Espectrometría de Masas en Tándem Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Ther Drug Monit Año: 2016 Tipo del documento: Article País de afiliación: Bélgica