Assessment of optical CT as a future QA tool for synchrotron x-ray microbeam therapy.
McErlean, Ciara M; Bräuer-Krisch, Elke; Adamovics, John; Doran, Simon J.
Afiliación
  • McErlean CM; CRUK Cancer Imaging Centre, Institute of Cancer Research, London, SM2 5NG, UK.
Phys Med Biol ; 61(1): 320-37, 2016 Jan 07.
Article en En | MEDLINE | ID: mdl-26657052
ABSTRACT
Synchrotron microbeam radiation therapy (MRT) is an advanced form of radiotherapy for which it is extremely difficult to provide adequate quality assurance. This may delay or limit its clinical uptake, particularly in the paediatric patient populations for whom it could be especially suitable. This study investigates the extent to which new developments in 3D dosimetry using optical computed tomography (CT) can visualise MRT dose distributions, and assesses what further developments are necessary before fully quantitative 3D measurements can be achieved. Two experiments are reported. In the first cylindrical samples of the radiochromic polymer PRESAGE(®) were irradiated with different complex MRT geometries including multiport treatments of collimated 'pencil' beams, interlaced microplanar arrays and a multiport treatment using an anthropomorphic head phantom. Samples were scanned using transmission optical CT. In the second experiment, optical CT measurements of the biologically important peak-to-valley dose ratio (PVDR) were compared with expected values from Monte Carlo simulations. The depth-of-field (DOF) of the optical CT system was characterised using a knife-edge method and the possibility of spatial resolution improvement through deconvolution of a measured point spread function (PSF) was investigated. 3D datasets from the first experiment revealed excellent visualisation of the 50 µm beams and various discrepancies from the planned delivery dose were found. The optical CT PVDR measurements were found to be consistently 30% of the expected Monte Carlo values and deconvolution of the microbeam profiles was found to lead to increased noise. The reason for the underestimation of the PVDR by optical CT was attributed to lack of spatial resolution, supported by the results of the DOF characterisation. Solutions are suggested for the outstanding challenges and the data are shown already to be useful in identifying potential treatment anomalies.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Terapia por Rayos X / Planificación de la Radioterapia Asistida por Computador / Tomografía Óptica Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2016 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Terapia por Rayos X / Planificación de la Radioterapia Asistida por Computador / Tomografía Óptica Límite: Humans Idioma: En Revista: Phys Med Biol Año: 2016 Tipo del documento: Article País de afiliación: Reino Unido