T2* Measurement bias due to concomitant gradient fields.
Hofstetter, Lorne W; Morrell, Glen; Kaggie, Joshua; Kim, Daniel; Carlston, Kristi; Lee, Vivian S.
Afiliación
  • Hofstetter LW; Department of Radiology, University of Utah, Salt Lake City, Utah, USA.
  • Morrell G; Department of Radiology, University of Utah, Salt Lake City, Utah, USA.
  • Kaggie J; Department of Radiology, University of Utah, Salt Lake City, Utah, USA.
  • Kim D; Department of Radiology, University of Cambridge, Cambridge, United Kingdom.
  • Carlston K; Department of Radiology, University of Utah, Salt Lake City, Utah, USA.
  • Lee VS; Department of Radiology, Northwestern University Feinberg School of Medicine, Chicago, Illinois, USA.
Magn Reson Med ; 77(4): 1562-1572, 2017 04.
Article en En | MEDLINE | ID: mdl-27186845
ABSTRACT

PURPOSE:

To demonstrate that concomitant magnetic fields can cause significant spatially dependent biases in T2* relaxometry measurements with implications for clinical applications such as BOLD and dynamic susceptibility contrast-enhanced MRI. THEORY AND

METHODS:

After developing a theoretical framework for intravoxel dephasing and signal loss from concomitant magnetic fields, this framework and the effect of concomitant fields on T2* are validated with phantom experiments and numerical simulation. In lower leg and renal T2* mapping, we quantify measurement bias for imaging protocols with high gradient amplitude multiecho readouts, comparable to those used in clinical applications.

RESULTS:

Concordance between phantom experiment and numerical simulation validate the theoretical framework. Changes in T2* measured in the lower leg and kidney varied by up to 15% and 35%, respectively, as a result of concomitant gradient effects when compared with the control measurements.

CONCLUSION:

Concomitant magnetic fields produced by imaging gradient coils can cause clinically significant T2* mapping errors when high amplitude, long duration gradient waveforms are used. While we have shown that measurement biases can be quite large, modification of imaging parameters can potentially reduce concomitant field-induced measurement errors to acceptable levels. Magn Reson Med 771562-1572, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Interpretación de Imagen Asistida por Computador / Artefactos / Músculo Esquelético Tipo de estudio: Diagnostic_studies / Guideline Límite: Female / Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética / Interpretación de Imagen Asistida por Computador / Artefactos / Músculo Esquelético Tipo de estudio: Diagnostic_studies / Guideline Límite: Female / Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos