Investigation of signal-to-noise ratio in frequency-domain multiphoton fluorescence lifetime imaging microscopy.
J Opt Soc Am A Opt Image Sci Vis ; 33(7): B1-B11, 2016 Jul 01.
Article en En | MEDLINE | ID: mdl-27409702
ABSTRACT
Multiphoton microscopy (MPM) combined with fluorescence lifetime imaging microscopy (FLIM) has enabled three-dimensional quantitative molecular microscopy in vivo. The signal-to-noise ratio (SNR), and thus the imaging rate of MPM-FLIM, which is fundamentally limited by the shot noise and fluorescence saturation, has not been quantitatively studied yet. In this paper, we investigate the SNR performance of the frequency-domain (FD) MPM-FLIM with two figures of merit the photon economy in the limit of shot noise, and the normalized SNR in the limit of saturation. The theoretical results and Monte Carlo simulations find that two-photon FD-FLIM requires 50% fewer photons to achieve the same SNR as conventional one-photon FLIM. We also analytically show that the MPM-FD-FLIM can exploit the DC and higher harmonic components generated by nonlinear optical mixing of the excitation light to improve SNR, reducing the required number of photons by an additional 50%. Finally, the effect of fluorophore saturation on the experimental SNR performance is discussed.

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: J Opt Soc Am A Opt Image Sci Vis Asunto de la revista: OFTALMOLOGIA Año: 2016 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: J Opt Soc Am A Opt Image Sci Vis Asunto de la revista: OFTALMOLOGIA Año: 2016 Tipo del documento: Article