FIELD CALIBRATION OF A TLD ALBEDO DOSEMETER IN THE HIGH-ENERGY NEUTRON FIELD OF CERF.
Radiat Prot Dosimetry
; 174(3): 315-321, 2017 Apr 28.
Article
en En
| MEDLINE
| ID: mdl-27421471
The new albedo dosemeter-type AWST-TL-GD 04 has been calibrated in the CERF neutron field (Cern-EU high-energy Reference Field). This type of albedo dosemeter is based on thermoluminescent detectors (TLDs) and used by the individual monitoring service of the Helmholtz Zentrum München (AWST) since 2015 for monitoring persons, who are exposed occupationally against photon and neutron radiation. The motivation for this experiment was to gain a field specific neutron correction factor Nn for workplaces at high-energy particle accelerators. Nn is a dimensionless factor relative to a basic detector calibration with 137Cs and is used to calculate the personal neutron dose in terms of Hp(10) from the neutron albedo signal. The results show that the sensitivity of the albedo dosemeter for this specific neutron field is not significantly lower as for fast neutrons of a radionuclide source like 252Cf. The neutron correction factor varies between 0.73 and 1.16 with a midrange value of 0.94. The albedo dosemeter is therefore appropriate to monitor persons, which are exposed at high-energy particle accelerators.