Gradient-echo EPI using a high-degree shim insert coil at 7 T: Implications for BOLD fMRI.
Kim, Tae; Lee, Yoojin; Zhao, Tiejun; Hetherington, Hoby P; Pan, Jullie W.
Afiliación
  • Kim T; Department of Radiology, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
  • Lee Y; Department of Bioengineering, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
  • Zhao T; Department of Radiology, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
  • Hetherington HP; Siemens Healthineers, USA, Siemens Medical Solution USA, Pittsburgh, Pennsylvania, USA.
  • Pan JW; Department of Radiology, University of Pittsburgh, Pittsburgh, Pennsylvania, USA.
Magn Reson Med ; 78(5): 1734-1745, 2017 11.
Article en En | MEDLINE | ID: mdl-27910126
PURPOSE: To quantitatively assess the effects of high degree and order (1st -4th+ ) relative to 1st -2nd degree B0 shimming at 7 Tesla (T) on gradient-echo echo planar imaging (GE-EPI) and blood-oxygen-level dependent (BOLD) activation. METHODS: Simulations and GE-EPI were performed at (2mm)3 and (3mm)3 resolution, evaluating the temporal signal-to-noise ratio (tSNR), transverse relaxivity ( R2*), BOLD % signal change and activated pixel counts in a breath-hold task. RESULTS: Comparing the 1st -4th+ degree with 1st -2nd degree shimmed B0 maps generated spatially varying regions of Δ|B0|=|B01-2|-|B01-4+|. As binned in 10-Hz intervals, the two center Δ|B0 | (±10 Hz) bins maintained the B0 offset of 48.6% of gray-matter pixels. In the positive Δ|B0 | bins greater than 10 Hz, the 1st -4th+ degree shimming improved the B0 offset in 41.1%; in negative Δ|B0 | bins less than -10 Hz, the offset worsened in 10.2% of the pixels. In the positive Δ|B0 | bins, we found variable but significant increases in BOLD sensitivity; the negative Δ|B0 | bins showed significant decreases. In the breath-hold studies, positive bins showed significantly increased activated pixel numbers (+5-29%), whereas negative bins showed -18 to 0% decline. CONCLUSION: 1st -4th+ degree shimming maintained B0 homogeneity over central brain regions while improving most of the other regions, including the inferior frontal lobe. Magn Reson Med 78:1734-1745, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Encéfalo / Imagen Eco-Planar Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Encéfalo / Imagen Eco-Planar Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos