The Effect of Pulsing on Transverse Ultrasound Efficiency and Chatter.
Wright, Alex J; Bohner, Austin D; Bernhisel, Ashlie A; Zaugg, Brian; Barlow, William R; Pettey, Jeff H; Olson, Randall J.
Afiliación
  • Wright AJ; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
  • Bohner AD; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
  • Bernhisel AA; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
  • Zaugg B; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
  • Barlow WR; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
  • Pettey JH; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
  • Olson RJ; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah. Electronic address: randallj.olson@hsc.utah.edu.
Am J Ophthalmol ; 183: 107-110, 2017 Nov.
Article en En | MEDLINE | ID: mdl-28860045
ABSTRACT

PURPOSE:

To evaluate the effects of micropulse, long pulse, and continuous ultrasound on transverse ultrasound using Abbott Medical Optics' (AMO) WhiteStar Signature Pro with the Ellips FX handpiece.

DESIGN:

In vitro laboratory study.

METHODS:

This study was conducted at the John A. Moran Eye Center Laboratory, University of Utah, Salt Lake City, Utah, USA. Porcine lenses were hardened in formalin for 2 hours and equilibrated in basic salt solution (BSS) over a 24-hour period. The lenses were then cubed in 2.0 × 2.0-mm pieces. These pieces were stored in BSS until the time of experimentation. The AMO WhiteStar Signature Pro machine (Abbott Medical Optics) with the Ellips FX handpiece and a 0.9-mm bent Dewey tip with a 30-degree bevel (Microsurgical Technology Inc) were used for phacoemulsification. Three runs of 20 lenses each were performed, measuring efficiency and chatter. Transverse ultrasound varied in the 3 runs and included continuous, 6 ms on/off micropulse, and 50 ms on/off long pulse.

RESULTS:

Micropulse was more efficient than long pulse by 43% (P = .00003) and continuous by 42% (P = .000387). There were also less chatter events with micropulse than with long-pulse and continuous ultrasound. However, this difference did not reach significance.

CONCLUSION:

The 6 ms on and 6 ms off micropulse transverse 3-dimensional ultrasound is more efficient and produces fewer chatter events than both long-pulse and continuous ultrasound.