Enhancement of the depth-of-field of integral imaging microscope by using switchable bifocal liquid-crystalline polymer micro lens array.
Opt Express ; 25(24): 30503-30512, 2017 Nov 27.
Article en En | MEDLINE | ID: mdl-29221078
ABSTRACT
An integral imaging microscopy (IIM) system with improved depth-of-field (DoF) using a custom-designed bifocal polarization-dependent liquid-crystalline polymer micro lens array (LCP-MLA) is proposed. The implemented MLA has improved electro-optical properties such as a small focal ratio, high fill factor, low driving voltage, and fast switching speed, utilizing a well-aligned reactive mesogen on the imprinted reverse shape of the lens and a polarization switching layer. A bifocal MLA switches its focal length according to the polarization angle and acquires different DoF information of the specimen. After two elemental image arrays are captured, the depth-slices are reconstructed and combined to provide a widened DoF. The fabricated bifocal MLA consists of two identical polarization-dependent LCP-MLAs with 1.6 mm and f/16 focal ratio. Our experimental results confirmed that the proposed system improves the DoF of IIM without the need for mechanical manipulation.

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: Opt Express Asunto de la revista: OFTALMOLOGIA Año: 2017 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: Opt Express Asunto de la revista: OFTALMOLOGIA Año: 2017 Tipo del documento: Article