Your browser doesn't support javascript.
loading
Intraoperative Shortening of Aneurysm Clips: Revisiting an Old Technique in a New Era.
Parmar, Vikas; Cikla, Ulas; Sayyahmelli, Sima; Ozaydin, Burak; Tumturk, Abdulfettah; Ahmed, Azam S; Baskaya, Mustafa K.
Afiliación
  • Parmar V; Department of Neurological Surgery University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
  • Cikla U; Department of Neurological Surgery University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
  • Sayyahmelli S; Department of Neurological Surgery University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
  • Ozaydin B; Department of Neurological Surgery University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
  • Tumturk A; Department of Neurological Surgery University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
  • Ahmed AS; Department of Neurological Surgery University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA.
  • Baskaya MK; Department of Neurological Surgery University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin, USA. Electronic address: baskaya@neurosurgery.wisc.edu.
World Neurosurg ; 139: 361-369, 2020 07.
Article en En | MEDLINE | ID: mdl-32339735
ABSTRACT

BACKGROUND:

It can sometimes be challenging to find a suitable clip to treat an unusual aneurysm, or when the surrounding anatomy is unusual, especially in resource-limited environments. We describe a method to modify aneurysm clips based on the method originated by Sugita et al in 1985. Herein clip modification (Clip-Mod) is used to treat anatomically difficult anterior communicating artery aneurysms.

METHODS:

The Department of Neurological Surgery database was reviewed to find aneurysm patients treated using modified aneurysm clips. Clip-Mod was performed during surgery by shortening the tines of titanium aneurysm clips by abrasion applied from the side of a standard 3-mm surgical diamond drill bit under constant irrigation. Note that the thickness of the tines and the clip spring were not modified or contacted by the drill.

RESULTS:

Four cases used modified aneurysm clips, from 648 total clip-treated aneurysms (0.6%) by 2 surgeons over a 14-year period. Three patients presented with subarachnoid hemorrhages that were determined to be due to anterior communicating artery aneurysms. One patient presented with an incidental unruptured anterior communicating artery aneurysm. All 4 patients were treated with 3-mm titanium clips shortened intraoperatively to 1- to 2-mm lengths, to achieve aneurysm obliteration without stenosing parent or perforating vessels.

CONCLUSIONS:

All 4 patients have done well clinically with no reoccurrences after 2-6 years' follow-up, which included angiographic evaluation. The use of this "Clip-Mod" technique thus appears useful for anterior communicating artery aneurysms. Clip-Mod could also be considered for treating other aneurysms when the "perfect" length clip is not available.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Hemorragia Subaracnoidea / Instrumentos Quirúrgicos / Aneurisma Intracraneal / Procedimientos Neuroquirúrgicos Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: World Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Hemorragia Subaracnoidea / Instrumentos Quirúrgicos / Aneurisma Intracraneal / Procedimientos Neuroquirúrgicos Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: World Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos