A Plasma Sample Preparation for Mass Spectrometry using an Automated Workstation.
Fu, Qin; Johnson, Casey W; Wijayawardena, Bhagya K; Kowalski, Michael P; Kheradmand, Miranda; Van Eyk, Jennifer E.
Afiliación
  • Fu Q; Advanced Clinical Biosystems Institute, Smidt Heart institute, Cedars Sinai Medical Center.
  • Johnson CW; Advanced Clinical Biosystems Institute, Smidt Heart institute, Cedars Sinai Medical Center.
  • Wijayawardena BK; Beckman Coulter Life Sciences.
  • Kowalski MP; Beckman Coulter Life Sciences.
  • Kheradmand M; Beckman Coulter Life Sciences.
  • Van Eyk JE; Advanced Clinical Biosystems Institute, Smidt Heart institute, Cedars Sinai Medical Center; jennifer.vaneyk@cshs.org.
J Vis Exp ; (158)2020 04 24.
Article en En | MEDLINE | ID: mdl-32391810
ABSTRACT
Sample preparation for mass spectrometry analysis in proteomics requires enzymatic cleavage of proteins into a peptide mixture. This process involves numerous incubation and liquid transfer steps in order to achieve denaturation, reduction, alkylation, and cleavage. Adapting this workflow onto an automated workstation can increase efficiency and reduce coefficients of variance, thereby providing more reliable data for statistical comparisons between sample types. We previously described an automated proteomic sample preparation workflow1. Here, we report the development of a more efficient and better controlled workflow with the following advantages 1) The number of liquid transfer steps is reduced from nine to six by combining reagents; 2) Pipetting time is reduced by selective tip pipetting using a 96-position pipetting head with multiple channels; 3) Potential throughput is increased by the availability of up to 45 deck positions; 4) Complete enclosure of the system provides improved temperature and environmental control and reduces the potential for contamination of samples or reagents; and 5) The addition of stable isotope labeled peptides, as well as ß-galactosidase protein, to each sample makes monitoring and quality control possible throughout the entire process. These hardware and process improvements provide good reproducibility and improve intra-assay and inter-assay precision (CV of less than 20%) for LC-MS based protein and peptide quantification. The entire workflow for digesting 96 samples in a 96-well plate can be completed in approximately 5 hours.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Espectrometría de Masas / Proteínas Sanguíneas / Proteómica / Métodos Analíticos de la Preparación de la Muestra Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Vis Exp Año: 2020 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Espectrometría de Masas / Proteínas Sanguíneas / Proteómica / Métodos Analíticos de la Preparación de la Muestra Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: J Vis Exp Año: 2020 Tipo del documento: Article