Your browser doesn't support javascript.
loading
Coated glass capillaries as SPME devices for DART mass spectrometry.
Cody, Robert; Maleknia, Simin D.
Afiliación
  • Cody R; JEOL USA, Inc, Peabody, MA, USA.
  • Maleknia SD; SDM Enterprises, Sydney, Australia.
Rapid Commun Mass Spectrom ; 34(23): e8946, 2020 Dec 15.
Article en En | MEDLINE | ID: mdl-32918514
ABSTRACT
RATIONALE Solid-phase microextraction (SPME) provides high-throughput sample cleanup and pre-concentration. Here we demonstrate coated glass capillaries (CGCs) as SPME devices for specific applications in direct analysis in real time (DART) mass spectrometry, referred to as "CGC-DART", for rapid screening of environmental contaminants at low parts-per-trillion detection limits and with accurate identification of analytes.

METHODS:

The extraction is performed in a one-step process in minutes by dipping the CGC in solutions containing the analytes, and then placing the CGC in a DART source for analysis. CGCs are disposable and relatively inexpensive in comparison with SPME devices, and can be prepared with hydrophobic, hydrophilic or mixed-mode materials similar to SPME. CGCs were prepared by adsorption coating with incubation of capillaries in saturated solutions of octadecylamine or covalent activation of silanes.

RESULTS:

Quantitation is shown with perfluorooctanoic acid (PFOA) at 1 ppt to 100 ppb, with the lowest detection at 500 parts-per quadrillion (ppq) in tap water. One-step extraction of contaminated groundwater from Northern Queensland, Australia, revealed perfluorooctane sulfonate (PFOS) and perfluorohexanesulfonamide as well as C4-C8 perfluoroalkyl carboxylic acids. A soil sample taken near a former military air base (New Hampshire, USA) revealed the presence of perfluorononanoic acid (PFNA) at 1 ppb and traces of perfluoroheptanoic acid.

CONCLUSIONS:

CGC-DART enabled one-step extraction of PFASs in minutes with mL sample volumes at low concentrations as shown for the standards and contaminated soil and water samples. DART-MS combined with Kendrick mass defect analysis enabled accurate identification of PFASs without chromatography steps, as fluorinated compounds are mass deficient and easily distinguished over background signal.

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: Rapid Commun Mass Spectrom Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: Rapid Commun Mass Spectrom Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos