Ultraviolet germicidal irradiation of the inner bore of a CT gantry.
Mahesh, Mahadevappa; Siewerdsen, Jeffrey H.
Afiliación
  • Mahesh M; The Russell H. Morgan Department of Radiology, Johns Hopkins University, Baltimore, MD, USA.
  • Siewerdsen JH; The Russell H. Morgan Department of Radiology, Johns Hopkins University, Baltimore, MD, USA.
J Appl Clin Med Phys ; 21(12): 325-328, 2020 Dec.
Article en En | MEDLINE | ID: mdl-33207042
PURPOSE: To investigate the feasibility and practicality of ultraviolet (UV) germicidal irradiation of the inner bore of a computed tomography (CT) gantry as a means of viral decontamination. METHOD: A UV lamp (PADNUT 38 W, 253 nm UV-C light tube) and UV-C dosimeter (GENERAL UV-C Digital Light Meter No. UV512C) were used to measure irradiance throughout the inner bore of a CT scanner gantry. Irradiance (units µW/cm2 ) was related to the time required to achieve 6-log viral kill (10-6 survival fraction). RESULTS: A warm-up time of ~120 s was required for the lamp to reach stable irradiance. Irradiance at the scan plane (z = 0 cm) of the CT scanner was 580.9 µW/cm2 , reducing to ~350 µW/cm2 at z = ±20 cm toward the front or back of the gantry. The angular distribution of irradiation was uniform within 10% coefficient of variation. A conservative estimate suggests at least 6-log kill (survival fraction ≤ 10-6 ) of viral RNA within ±20 cm of the scan plane with an irradiation time of 120 s from cold start. More conservatively, running the lamp for 180 s (3 min) or 300 s (5 min) from cold start is estimated to yield survival fraction <<10-7 survival fraction within ±20 cm of the scan plane. CONCLUSION: Ultraviolet irradiation of the inner bore of the CT gantry can be achieved with a simple UV-C lamp attached to the CT couch. Such practice could augment manual wipe-down procedures, improve safety for CT technologists or housekeeping staff, and could potentially reduce turnover time between scanning sessions.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Tomógrafos Computarizados por Rayos X / Desinfección / Control de Infecciones / COVID-19 Tipo de estudio: Diagnostic_studies / Guideline Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Tomógrafos Computarizados por Rayos X / Desinfección / Control de Infecciones / COVID-19 Tipo de estudio: Diagnostic_studies / Guideline Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos