Radiohistologic Comparison Study of Temporal Bone Specimens After Cochlear Implant Electrode Array Insertion: Is Cone-Beam CT Superior to MDCT?
Burck, Iris; Schneider, Serena V; Balster, Sven; Lehn, Annette; Yel, Ibrahim; Albrecht, Moritz H; Helbig, Silke; Stöver, Timo; Kaltenbach, Benjamin; Vogl, Thomas J.
Afiliación
  • Burck I; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
  • Schneider SV; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
  • Balster S; Department of Otorhinolaryngology, University Hospital Frankfurt, Frankfurt, Germany.
  • Lehn A; Institute of Biostatistics and Mathematical Modeling, University Hospital Frankfurt, Frankfurt, Germany.
  • Yel I; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
  • Albrecht MH; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
  • Helbig S; Department of Otorhinolaryngology, University Hospital Frankfurt, Frankfurt, Germany.
  • Stöver T; Department of Otorhinolaryngology, University Hospital Frankfurt, Frankfurt, Germany.
  • Kaltenbach B; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
  • Vogl TJ; Department of Diagnostic and Interventional Radiology, University Hospital Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
AJR Am J Roentgenol ; 216(3): 752-758, 2021 03.
Article en En | MEDLINE | ID: mdl-33439050
ABSTRACT
OBJECTIVE. The purpose of this article is to evaluate subjective image quality and diagnostic accuracy to determine cochlear implant (CI) electrode position in a temporal bone (TB) specimen on cone-beam CT (CBCT) versus MDCT. MATERIALS AND METHODS. In this retrospective study, two radiologists independently reviewed CBCT (96-kV and 120-kV settings) and MDCT images of 20 TB specimens after electrode implantation. Qualitative evaluation of bone structures of the otic capsule, inner and outer cochlear wall, osseous spiral lamina, electrode position relative to the osseous spiral lamina, visualization of single electrode contacts on the array, metal artifacts, and overall image quality was performed using a five-point scale. Intracochlear electrode position was subsequently correlated with histologic examination. RESULTS. Radiologic assessment of bone structures of the otic capsule, the cochlear wall (except the outer part), osseous spiral lamina, electrode position, visualization of single electrode contacts on the array, metal artifacts, and overall image quality were significantly higher in CBCT compared with MDCT (maximum p = .04). No significant differences were found between CBCT at 96 kV and 120 kV (minimum p = .21). The intracochlear electrode position with histologic correlation was correctly diagnosed in 100% and 97.5% of specimens on 120-kV and 96-kV CBCT, respectively, whereas 77.5% were correctly assessed using MDCT. CONCLUSION. The data suggest that CBCT shows a higher diagnostic accuracy in TB specimen imaging after CI compared with MDCT, in particular to determine the intracochlear localization of the implant.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Hueso Temporal / Implantes Cocleares / Implantación Coclear / Tomografía Computarizada de Haz Cónico / Tomografía Computarizada Multidetector Tipo de estudio: Observational_studies / Qualitative_research Límite: Humans Idioma: En Revista: AJR Am J Roentgenol Año: 2021 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Hueso Temporal / Implantes Cocleares / Implantación Coclear / Tomografía Computarizada de Haz Cónico / Tomografía Computarizada Multidetector Tipo de estudio: Observational_studies / Qualitative_research Límite: Humans Idioma: En Revista: AJR Am J Roentgenol Año: 2021 Tipo del documento: Article País de afiliación: Alemania