Comparison of ablation performance between dual internally cooled wet tip and conventional dual internally cooled tip radiofrequency electrodes: an experimental study in ex vivo bovine liver.
Cha, Dong Ik; Lee, Min Woo; Jeong, Woo Kyoung; Ha, Sang Yun; Ahn, Soo Hyun; Rhim, Hyunchul; Lim, Hyo Keun.
Afiliación
  • Cha DI; Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Lee MW; Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Jeong WK; Department of Health Sciences and Technology, SAIHST, Sungkyunkwan University, Seoul, Republic of Korea.
  • Ha SY; Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Ahn SH; Department of Health Sciences and Technology, SAIHST, Sungkyunkwan University, Seoul, Republic of Korea.
  • Rhim H; Department of Pathology and Translational Genomics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Lim HK; Department of Mathematics, Ajou University, Suwon, Republic of Korea.
Int J Hyperthermia ; 38(1): 332-340, 2021.
Article en En | MEDLINE | ID: mdl-33627017
ABSTRACT

OBJECTIVE:

To evaluate the performance of dual internally cooled wet tip (ICWT) radiofrequency electrodes in comparison to dual internally cooled tip (ICT) electrodes.

METHODS:

Twenty ablation zones were created for each type of electrodes. Planned procedure time was 6 min. Diameters of the ablation zone along the x-, y-, and z-axes (Dx, Dy, and Dz), ablation zone sphericity, quantitative sphericity measurement, and ablation volume were measured and compared between the two electrode types. Circularity of the ablation zone on the surface with x- and z- axes (zx plane) and amount of energy applied were also compared.

RESULTS:

Dx and Dz were significantly longer with ICWT than those with ICT (Dx 3.0 vs. 2.8 cm, p = .018; and Dz 2.7 vs. 2.3 cm, p < .001, respectively). Dy was not significantly different (3.0 vs. 2.9 cm, p = .220). Moreover, 85% (17/20) and 30% (6/20) of ablation zones from ICWT and ICT were spherical (p = .001), respectively. Quantitative measurement showed that ICWT was more spherical compared to ICT (0.962 vs. 0.881, p = .001). The ablation volume was also significantly higher with ICWT (11.55 vs. 9.45 cm3, p = .003). The ablation zone on the zx plane was more circular with ICWT (0.907 vs. 0.883, p = .028). The amount of energy applied was significantly bigger with ICWT (18508 vs. 16998 WS, p = .003).

CONCLUSION:

Dual ICWT electrodes were better able to create more spherical and larger ablation zones than dual ICT electrodes.