Sensitive, wide-range and high-throughput quantification of cyclosporine in whole blood using ultra-performance liquid chromatography coupled to tandem mass spectrometry and comparison with an antibody-conjugated magnetic immunoassay.
Watanabe, Takuma; Tanaka, Ryota; Ono, Hiroyuki; Suzuki, Yosuke; Tatsuta, Ryosuke; Itoh, Hiroki.
Afiliación
  • Watanabe T; Department of Clinical Pharmacy, Oita University Hospital, Yufu, Oita, Japan.
  • Tanaka R; Department of Clinical Pharmacy, Oita University Hospital, Yufu, Oita, Japan.
  • Ono H; Department of Clinical Pharmacy, Oita University Hospital, Yufu, Oita, Japan.
  • Suzuki Y; Department of Medication Use Analysis and Clinical Research, Meiji Pharmaceutical University, Tokyo, Japan.
  • Tatsuta R; Department of Clinical Pharmacy, Oita University Hospital, Yufu, Oita, Japan.
  • Itoh H; Department of Clinical Pharmacy, Oita University Hospital, Yufu, Oita, Japan.
Biomed Chromatogr ; 35(8): e5128, 2021 Aug.
Article en En | MEDLINE | ID: mdl-33780006
ABSTRACT
Because either trough or peak concentration at 2 h after administration is measured in routine therapeutic drug monitoring for cyclosporine A (CyA), a quantification method with a wide-range calibration curve capable of simultaneously measuring both concentrations is required. We developed a sensitive, wide-range and high-throughput quantification method for CyA in whole blood using ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS), and compared patients' blood CyA levels measured by UPLC-MS/MS and antibody-conjugated magnetic immunoassay (ACMIA). Whole blood samples were prepared by solid-phase extraction using Oasis HLB µElution plate. The UPLC-MS/MS assay showed excellent linearity over a wide calibration range of 5-2500 ng/mL. Within-batch accuracy and precision as well as batch-to-batch accuracy and precision fulfilled the criteria of US Food and Drug Administration guidelines. The blood CyA concentrations measured by the UPLC-MS/MS assay correlated strongly with those measured by ACMIA. A Bland-Altman plot showed a fixed error between CyA concentrations measured by the two methods, and the concentrations measured by the UPLC-MS/MS method were consistently lower than those measured by ACMIA. We have succeeded to develop a sensitive, wide-range and high-throughput quantification method for CyA in whole blood using UPLC-MS/MS.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Inmunoensayo / Cromatografía Líquida de Alta Presión / Ciclosporina / Espectrometría de Masas en Tándem Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Biomed Chromatogr Año: 2021 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Inmunoensayo / Cromatografía Líquida de Alta Presión / Ciclosporina / Espectrometría de Masas en Tándem Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Biomed Chromatogr Año: 2021 Tipo del documento: Article País de afiliación: Japón