Transitional conic toric intraocular lens evaluation after femtosecond laser-assisted cataract surgery using intraoperative aberrometry.
Pastor-Pascual, Francisco; Pastor-Pascual, Rafael; Montés-Micó, Robert; Ruiz-Mesa, Ramón; Tañá-Rivero, Pedro.
Afiliación
  • Pastor-Pascual F; Oftalvist, C/Ruzafa 19, 46004, Valencia, Spain. franciscopastorpascual@gmail.com.
  • Pastor-Pascual R; Oftalvist, C/Ruzafa 19, 46004, Valencia, Spain.
  • Montés-Micó R; Oftalvist, C/Ruzafa 19, 46004, Valencia, Spain.
  • Ruiz-Mesa R; University of Valencia, Valencia, Spain.
  • Tañá-Rivero P; Oftalvist, C/Ruzafa 19, 46004, Valencia, Spain.
Int Ophthalmol ; 42(1): 177-189, 2022 Jan.
Article en En | MEDLINE | ID: mdl-34424437
ABSTRACT

PURPOSE:

To assess refractive and visual outcomes following phacoemulsification with femtosecond laser-assisted cataract surgery (FLACS) using intraoperative aberrometry and implantation of a toric intraocular lens (IOL) in eyes with different degrees of astigmatism.

METHODS:

One hundred two eyes of 70 patients who underwent implantation of the transitional toric 565 Precizon IOL (Ophtec BV) were enrolled. FLACS, capsular tension ring insertion, and intraoperative aberrometry were performed. Main outcome measures were refractive error, uncorrected- and corrected distance snellen decimal visual acuity values (UDVA and CDVA, respectively), and IOL rotation. Specifically, a vector analysis was carried out with J0 and J45 evaluation. Eyes were evaluated 1-year after surgery.

RESULTS:

Overall, 94.12% (96 eyes) and 100% (102 eyes) of the eyes showed a spherical equivalent (SE) within ± 0.50D and ± 1.00D, respectively. The mean SE and refractive cylinder were - 0.06 ± 0.29D and - 0.23 ± 0.37D, respectively. Vector analysis revealed that 100% of the eyes were within ± 0.50D for the J0 and J45 cylindrical components. The mean toric axis rotation was 1.10 ± 1.71° (from 0° to 5°), 77% (79 eyes), and 100% (102 eyes) of the eyes showed UDVA and CDVA of 20/25, respectively. The postoperative mean values of monocular UDVA and CDVA were 0.88 ± 0.17 and 0.96 ± 0.07 (about 20/20), respectively. No patient required IOL realignment during the postoperative follow-up.

CONCLUSIONS:

The present study suggests that the use of the Precizon IOL after FLACS, using intraoperative aberrometry in patients with different amounts of astigmatism, provides good visual acuity, accurate refractive outcomes, and excellent rotational stability.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Astigmatismo / Catarata / Facoemulsificación / Lentes Intraoculares Tipo de estudio: Diagnostic_studies / Etiology_studies / Observational_studies Límite: Humans Idioma: En Revista: Int Ophthalmol Año: 2022 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Astigmatismo / Catarata / Facoemulsificación / Lentes Intraoculares Tipo de estudio: Diagnostic_studies / Etiology_studies / Observational_studies Límite: Humans Idioma: En Revista: Int Ophthalmol Año: 2022 Tipo del documento: Article País de afiliación: España