Your browser doesn't support javascript.
loading
Characterization and commissioning of a Leksell Gamma Knife ICON system for framed and frameless stereotactic radiosurgery.
Hu, Yue-Houng; Hickling, Susannah V; Qian, Jing; Blackwell, C Robert; McLemore, Luke B; Tryggestad, Erik J.
Afiliación
  • Hu YH; Department of Radiation Oncology, Division of Medical Physics, Mayo Clinic, Rochester, Minnesota.
  • Hickling SV; Department of Radiation Oncology, Division of Medical Physics, Mayo Clinic, Rochester, Minnesota.
  • Qian J; Department of Radiation Oncology, Division of Medical Physics, Mayo Clinic, Rochester, Minnesota.
  • Blackwell CR; Department of Radiation Oncology, Division of Medical Physics, Mayo Clinic, Rochester, Minnesota.
  • McLemore LB; Department of Radiation Oncology, Division of Medical Physics, Mayo Clinic, Rochester, Minnesota.
  • Tryggestad EJ; Department of Radiation Oncology, Division of Medical Physics, Mayo Clinic, Rochester, Minnesota.
J Appl Clin Med Phys ; 23(3): e13475, 2022 Mar.
Article en En | MEDLINE | ID: mdl-35064749
ABSTRACT

PURPOSE:

The Leksell Gamma Knife Icon unit (IU) was introduced recently as an upgrade to the Perfexion unit (PU) at our Gamma Knife practice. In the current study, we sought mainly to characterize dosimetry and targeting accuracy of the IU treatment deliveries using both invasive frame and frameless treatment workflows.

METHODS:

Relative output factors were measured by delivering single-shot 4, 8 and 16 mm radiation profiles in the manufacturer's acrylonitrile butadiene styrene spherical phantom in coronal and sagittal planes using EBT3 film. Resultant dosimetry was compared with the manufacturer's dose calculation and derived output factors were compared with the manufacturer's published value. Geometric consistency of stereotactic coordinates based on cone-beam computed tomography (CBCT) versus the traditional conventional CT-based method was characterized using a rigid phantom containing nine fiducial indicators over four separate trials. End-to-end (E2E) testing using EBT3 film was designed to evaluate both dosimetric and geometric accuracy for hypothetical framed and frameless workflows.

RESULTS:

Relative output factors as measured by the manufacturer were independently confirmed using EBT3 film measurements to within 2%. The mean 3D radial discrepancy in stereotactic space between CBCT and CT-based definition over the sampled locations in our rigid geometry phantom was demonstrated to be between 0.40 mm and 0.56 mm over the set of trials, larger than prior reported values. E2E performed in 2D demonstrates sub-mm (and typically < 0.5 mm) accuracy for framed and frameless workflows; geometric accuracy of framed treatments using CBCT-defined stereotactic coordinates was shown to be slightly improved in comparison with those defined using conventional CT. Furthermore, in phantom, frameless workflows exhibited better accuracy than framed workflows for fractionated treatments, despite large magnitudes of introduced interfraction setup error. Accuracy of dosimetric delivery was confirmed in terms of qualitative comparisons of dose profiles and in terms of 2D gamma pass rates based on 1%/1 mm criteria.

CONCLUSION:

The IU was commissioned for clinical use of frameless and framed treatment protocols. The present study outlines an extensive E2E methodology for confirmation of dosimetric and geometric treatment accuracy.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Radiocirugia Tipo de estudio: Guideline / Prognostic_studies / Qualitative_research Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2022 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Radiocirugia Tipo de estudio: Guideline / Prognostic_studies / Qualitative_research Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2022 Tipo del documento: Article