Your browser doesn't support javascript.
loading
De Ritis-adjusted AST provides comparable efficacy to lactate dehydrogenase as a biomarker for detection of LVAD hemolysis or thrombosis.
Schettle, Sarah; Rosenbaum, Andrew; Goodman, Daniel; Stulak, John.
Afiliación
  • Schettle S; Cardiovascular Surgery, Mayo Clinic, Rochester, Minnesota, USA.
  • Rosenbaum A; Advanced Heart Failure Cardiology and Cardiac Transplant, Mayo Clinic, Rochester, Minnesota, USA.
  • Goodman D; Cardiovascular Surgery, Mayo Clinic, Rochester, Minnesota, USA.
  • Stulak J; Cardiovascular Surgery, Mayo Clinic, Rochester, Minnesota, USA.
Artif Organs ; 46(7): 1425-1428, 2022 Jul.
Article en En | MEDLINE | ID: mdl-35318681
ABSTRACT

BACKGROUND:

Lactate hydrogenase (LDH) is a common biomarker utilized in the detection and monitoring of left ventricular assist device (LVAD) hemolysis and thrombosis. Aspartate aminotransferase (AST) and alanine aminotransferase (ALT) are common laboratory tests that can be used to calculate the De Ritis ratio and the De Ritis adjusted AST.

METHODS:

A retrospective review of LVAD patients was performed to identify three cohorts of patients those with confirmed pump thrombosis after device exchange, those with LVAD-related hemolysis who were medically managed without pump exchange, and those who did not meet these criteria and served as the control cohort. Evaluation of AST, AST/ALT ratio (referred to as the De Ritis ratio) as well as AST x (AST/ALT) or the De Ritis-adjusted AST (DRA) was performed.

RESULTS:

There were 29 patients who underwent device exchange for thrombosis, 25 patients who were diagnosed with hemolysis and treated medically (clopidogrel (N = 6), heparin (N = 13), tirofiban (N = 8), eptifibatide (N = 2), and some received more than one of these treatments), and 425 control patients. A qualitatively comparable relative and absolute rise in DRA and LDH were found in both surgically managed pump thrombosis and suspected device-related hemolysis.

CONCLUSIONS:

Both AST and LDH as well DRA are significantly associated with pump thrombosis (p < 0.0001 for each). DRA is a potential screening biomarker for hemolysis and device thrombosis in stable left ventricular assist device patients.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Trombosis / Corazón Auxiliar Tipo de estudio: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Artif Organs Año: 2022 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Trombosis / Corazón Auxiliar Tipo de estudio: Diagnostic_studies / Etiology_studies / Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Artif Organs Año: 2022 Tipo del documento: Article País de afiliación: Estados Unidos