First-generation clinical dual-source photon-counting CT: ultra-low-dose quantitative spectral imaging.
Eur Radiol
; 32(12): 8579-8587, 2022 Dec.
Article
en En
| MEDLINE
| ID: mdl-35708838
OBJECTIVE: Evaluation of image characteristics at ultra-low radiation dose levels of a first-generation dual-source photon-counting computed tomography (PCCT) compared to a dual-source dual-energy CT (DECT) scanner. METHODS: A multi-energy CT phantom was imaged with and without an extension ring on both scanners over a range of radiation dose levels (CTDIvol 0.4-15.0 mGy). Scans were performed in different modes of acquisition for PCCT with 120 kVp and DECT with 70/Sn150 kVp and 100/Sn150 kVp. Various tissue inserts were used to characterize the precision and repeatability of Hounsfield units (HUs) on virtual mono-energetic images between 40 and 190 keV. Image noise was additionally investigated at an ultra-low radiation dose to illustrate PCCT's ability to remove electronic background noise. RESULTS: Our results demonstrate the high precision of HU measurements for a wide range of inserts and radiation exposure levels with PCCT. We report high performance for both scanners across a wide range of radiation exposure levels, with PCCT outperforming at low exposures compared to DECT. PCCT scans at the lowest radiation exposures illustrate significant reduction in electronic background noise, with a mean percent reduction of 74% (p value ~ 10-8) compared to DECT 70/Sn150 kVp and 60% (p value ~ 10-6) compared to DECT 100/Sn150 kVp. CONCLUSIONS: This paper reports the first experiences with a clinical dual-source PCCT. PCCT provides reliable HUs without disruption from electronic background noise for a wide range of dose values. Diagnostic benefits are not only for quantification at an ultra-low dose but also for imaging of obese patients. KEY POINTS: PCCT scanners provide precise and reliable Hounsfield units at ultra-low dose levels. The influence of electronic background noise can be removed at ultra-low-dose acquisitions with PCCT. Both spectral platforms have high performance along a wide range of radiation exposure levels, with PCCT outperforming at low radiation exposures.
Palabras clave
Texto completo:
1
Bases de datos:
MEDLINE
Asunto principal:
Tomografía Computarizada por Rayos X
Límite:
Humans
Idioma:
En
Revista:
Eur Radiol
Asunto de la revista:
RADIOLOGIA
Año:
2022
Tipo del documento:
Article
País de afiliación:
Estados Unidos