Use of tandem carbon microfiber columns for on-line fractionation strategy of reducing ion suppression effects in electrospray ionization-mass spectrometry.
Wang, Zhao; Shang, Hai-Bo; Cui, Meiyu; Ma, Biao; Li, Donghao; Jin, Dongri; Dong, Meihua.
Afiliación
  • Wang Z; Department of Chemistry, Yanbian University, Yanji City, Jilin Province 133002, P.R. China.
  • Shang HB; Department of Chemistry, Yanbian University, Yanji City, Jilin Province 133002, P.R. China; Interdisciplinary Program of Biological Functional Molecules, College of Integration Science, Yanbian University, Park Road 977, Yanji City, Jilin Province 133002, PR China. Electronic address: hbshang@ybu.ed
  • Cui M; Department of Chemistry, Yanbian University, Yanji City, Jilin Province 133002, P.R. China.
  • Ma B; College of Chemistry, State Key Laboratory of Medicinal Chemical Biology, Nankai University, Tianjin 300071, PR China.
  • Li D; Department of Chemistry, Yanbian University, Yanji City, Jilin Province 133002, P.R. China.
  • Jin D; Department of Chemistry, Yanbian University, Yanji City, Jilin Province 133002, P.R. China. Electronic address: drjin@ybu.edu.cn.
  • Dong M; Department of Chemistry, Yanbian University, Yanji City, Jilin Province 133002, P.R. China. Electronic address: mhdong@ybu.edu.cn.
J Chromatogr A ; 1685: 463641, 2022 Dec 06.
Article en En | MEDLINE | ID: mdl-36403517
ABSTRACT
As we all know, the complexity and diversity of complex sample are confronting with challenge of high-sensitive mass spectrometry analysis, especially direct mass spectrometry. The work proposed a two-dimensional carbon microfiber fractionation (2DµCFs) system for the reduction of ion suppression effects in electrospray ionization mass spectrometry (ESI-MS). The 2DµCFs system can on-line fractionated the complex sample into strong-polar, medium-polar and weak-polar fractions for sequential MS analysis. Direct analysis brings about the strong ion suppression effect up to 85%, but the fractionated analysis of 2DµCFs system can distinctly reduce the ion suppression effect to less than 43%, even close to none. And the fractionated analysis not only decrease the number of analytes of direct analysis, but also narrows down the polarity range of analytes within the droplets of ESI, contributing to the homogeneous distribution to reduce the ion suppression effect. As an example, the 2DµCFs system coupled with tandem mass spectrometry (MS/MS) was applied for fractionated analysis of Radix Puerariae extract in 4.5 min. Compared with direct MS/MS, the 2DµCFs-MS/MS shows the lower ion suppression and the more ionic species (m/z). In addition, and most of ionic species detected by the 2DµCFs-MS/MS, are the same as those by HPLC MS/MS. Furthermore, the 2DµCFs-MS/MS exhibit the good analysis repeatability of real sample with the RSDs less than 10.32% (intra-day), 7.12% (inter-day) and 14.28% (inter-batch of CFs and ACFs). The carbon fibers (CFs) and active carbon fibers (ACFs) columns, as the key parts, are conducive to achieve on-line fractionation of compounds based on the difference of polarity. The 2DµCFs system has the merits of on-line, speediness, low-pressure and recycle. More importantly, such fast and high-throughput method is advantageous for comprehensive screening of complex samples in drug, clinical, environment and plant.