Performance of a new auxiliary device based on wrist brace to improve accuracy and feasibility in neuromuscular monitoring with acceleromyography in prone-positioned patients undergoing lumbar spine surgery: a prospective randomized clinical trial.
J Clin Monit Comput
; 37(4): 993-1001, 2023 08.
Article
en En
| MEDLINE
| ID: mdl-37004664
Accuracy of acceleromyography (AMG) is not be comparable to that of mechanomyography or electromyography (EMG). In particular, the prone position may reduce the accuracy and feasibility of AMG. We developed a new device based on wrist brace to allow free thumb movement and fix the other parts of the hand and wrist. We aimed to test whether the brace applied to the AMG would increase precision of AMG and agreement with the EMG in the prone position. Fifty-seven patients, undergoing lumbar surgery under general anesthesia, were randomly assigned to groups to which AMG was applied with or without (no) brace (29 in group B vs. 28 in group NB). EMG was performed in the contralateral arm. Repeatability coefficients of the first twitch height (T1) and train-of-four (TOF) ratio were assessed from nine consecutive measurements during spontaneous recovery from rocuronium-induced neuromuscular block and the AMGs of the two groups were compared in prone position. The agreement between AMG and EMG in each group was assessed using the Bland-Altman method. In group B, the repeatability coefficient of T1 was significantly lower during the recovery to T1 of 25% and TOF ratio of 0.9 (P = 0.017 and 0.033, respectively), indicating higher precision. The mean differences of bias (95% limits of agreement) between AMG and EMG in TOF ratio of 0.9 were 6.839 (- 26.54 to 40.22) in group NB and 3.922 (- 21.83 to 29.67) in group B. The wide limits of agreement in group NB was slightly narrowed in group B but without significance. Trial registration: registered on the UMIN Clinical Trials Registry in August 2020 (UMIN000041310).
Palabras clave
Texto completo:
1
Bases de datos:
MEDLINE
Asunto principal:
Bloqueo Neuromuscular
/
Monitoreo Neuromuscular
Tipo de estudio:
Clinical_trials
/
Observational_studies
Límite:
Humans
Idioma:
En
Revista:
J Clin Monit Comput
Asunto de la revista:
INFORMATICA MEDICA
/
MEDICINA
Año:
2023
Tipo del documento:
Article