HS-GC-FID method for quantification of HFA-152a in cell culture media, and plasma from a range of species and regulatory compliant validations.
Kuehl, Philip J; Corr, Stuart; Farrar, Jabari; McDonald, Jacob D; Wermer, Tim; Weber, Derek; Leach, Chet.
Afiliación
  • Kuehl PJ; Lovelace Biomedical, 2425 Ridgecrest Dr. SE, Albuquerque, NM 87108, United States of America. Electronic address: pkuehl@lovelacebiomedical.org.
  • Corr S; Koura, Thornton Science Park, Cheshire, UK.
  • Farrar J; Lovelace Biomedical, 2425 Ridgecrest Dr. SE, Albuquerque, NM 87108, United States of America.
  • McDonald JD; Lovelace Biomedical, 2425 Ridgecrest Dr. SE, Albuquerque, NM 87108, United States of America.
  • Wermer T; Lovelace Biomedical, 2425 Ridgecrest Dr. SE, Albuquerque, NM 87108, United States of America.
  • Weber D; Lovelace Biomedical, 2425 Ridgecrest Dr. SE, Albuquerque, NM 87108, United States of America.
  • Leach C; Lovelace Biomedical, 2425 Ridgecrest Dr. SE, Albuquerque, NM 87108, United States of America.
J Pharmacol Toxicol Methods ; 122: 107271, 2023.
Article en En | MEDLINE | ID: mdl-37196729
ABSTRACT

INTRODUCTION:

1,1-Difluoroethane (HFA-152a) is being developed as an alternative propellant in pressurized metered dose inhalers (pMDIs). As a part of the regulatory development pathway, pharmacology, toxicology and clinical studies have been conducted with inhaled HFA-152a. These studies require fit for purpose regulatory compliant (GxP validated) methods for quantification of HFA-152a from blood.

METHODS:

As HFA-152a is a gas at standard temperature and pressure, novel methods were developed to support the analysis across the wide range of species and concentrations required for regulatory filing.

RESULTS:

The developed methods utilized a headspace auto sampler coupled to a gas chromatograph (GC) with flame ionization detection. Key factors in the successful method included bringing together fit for purpose approaches to the head space vials, volume of matrix (blood), detection range required for species/study objective, handling / transfer of blood into head space vials and the stability/storage required for the analysis of the samples. The species-specific assays were fully validated under regulatory (GLP) conditions for mouse, rat, rabbit, canine and human and non-regulatory (non GLP) validations for guinea pig and cell culture media.

DISCUSSION:

Overall the novel approach of head space analysis of whole blood allowed for the development and validation of assays used to generate the toxicokinetic data that supported clinical testing of HFA-152a as a new pMDI propellant.