Magn Reson Med
; 91(6): 2508-2518, 2024 Jun.
Article
en En
| MEDLINE
| ID: mdl-38321602
ABSTRACT
PURPOSE:
The purpose of this study is to improve the image quality of diffusion-weighted images obtained with a single RF transmit channel 7 T MRI setup using time-resampled frequency-offset corrected inversion (TR-FOCI) pulses to refocus the spins in a twice-refocused spin-echo readout scheme.METHODS:
We replaced the conventional Shinnar-Le Roux-pulses in the twice refocused diffusion sequence with TR-FOCI pulses. The slice profiles were evaluated in simulation and experimentally in phantoms. The image quality was evaluated in vivo comparing the Shinnar-Le Roux and TR-FOCI implementation using a b value of 0 and of 1000 s/mm2.RESULTS:
The b0 and diffusion-weighted images acquired using the modified sequence improved the image quality across the whole brain. A region of interest-based analysis showed an SNR increase of 113% and 66% for the nondiffusion-weighted (b0) and the diffusion-weighted (b = 1000 s/mm2) images in the temporal lobes, respectively. Investigation of all slices showed that the adiabatic pulses mitigated B 1 + $$ {B}_1^{+} $$ inhomogeneity globally using a conventional single-channel transmission setup.CONCLUSION:
The TR-FOCI pulse can be used in a twice-refocused spin-echo diffusion pulse sequence to mitigate the impact of B 1 + $$ {B}_1^{+} $$ inhomogeneity on the signal intensity across the brain at 7 T. However, further work is needed to address SAR limitations.Palabras clave
Texto completo:
1
Bases de datos:
MEDLINE
Asunto principal:
Algoritmos
/
Imagen de Difusión por Resonancia Magnética
Idioma:
En
Revista:
Magn Reson Med
/
Magn. Reson. Med
/
Magnetic Resonance in Medicine
Asunto de la revista:
DIAGNOSTICO POR IMAGEM
Año:
2024
Tipo del documento:
Article
País de afiliación:
Australia