Your browser doesn't support javascript.
loading
Patient-Specific Commissural Alignment With a Self-Expanding Transcatheter Heart Valve.
Cruz-González, Ignacio; Antúnez-Muiños, Pablo; López-Tejero, Sergio; Arnold, Lars; Oldroyd, Keith G.
Afiliación
  • Cruz-González I; Cardiology Department, Hospital Universitario de Salamanca, Salamanca, Spain.
  • Antúnez-Muiños P; Biomedical Research Institute of Salamanca, Salamanca, Spain.
  • López-Tejero S; Centro de Investigación Biomédica En Red. Enfermedades Cardiovasculares (CIBER-CV), Salamanca, Spain.
  • Arnold L; Cardiology Department, Hospital Universitario de Salamanca, Salamanca, Spain.
  • Oldroyd KG; Biomedical Research Institute of Salamanca, Salamanca, Spain.
JACC Case Rep ; 29(4): 102190, 2024 Feb 21.
Article en En | MEDLINE | ID: mdl-38379652
ABSTRACT
A simple and reproducible technique to achieve commissural alignment during transcatheter aortic valve replacement with the Allegra valve is described. Slight rotation of the system before system insertion is necessary. Moreover, thanks to its permaflow system (Biosensors) and its radiopaque markings, small adjustments before valve deployment can be made to reassess correct alignment.
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: JACC Case Rep Año: 2024 Tipo del documento: Article País de afiliación: España