Visual and Refractive Outcomes After Bi-aspheric Trifocal Toric Diffractive Intraocular Lens Implantation.
J Refract Surg ; 40(6): e407-e419, 2024 May.
Article en En | MEDLINE | ID: mdl-38848051
ABSTRACT

PURPOSE:

To assess clinical outcomes after cataract surgery with bilateral implantation of a new bi-aspheric diffractive intraocular lens (IOL).

METHODS:

Thirty patients underwent bilateral implantation of the Asqelio Trifocal Toric IOL (AST Products, Inc) and were evaluated 3 months postoperatively. Main outcomes included refractive error, photopic monocular and binocular uncorrected and corrected distance (UDVA, CDVA), intermediate (UIVA, CDIVA) at 60 cm, and near (UNVA, CDNVA) at 40 cm visual acuities. Mesopic monocular and binocular CDNVA were also measured. Defocus curves, binocular contrast sensitivity under photopic and mesopic conditions with and without glare and rotational stability were determined. Patients completed Catquest-9SF and visual symptoms questionnaires.

RESULTS:

Average values of binocular photopic CDVA, CDIVA, and CDNVA, and mesopic CDNVA were -0.04 ± 0.06, 0.02 ± 0.08, 0.02 ± 0.07, and 0.22 ± 0.11 logMAR, respectively. All patients achieved cumulative CDVA ⩾ 20/25, and CDIVA and CDNVA of 20/32 or better. Binocular depth of focus was approximately 3.25 diopters (D). Mean postoperative spherical equivalent was -0.08 ± 0.26 D, with 95% of eyes within ±0.50 D. Mean postoperative refractive cylinder was -0.22 ± 0.27 D, with 91.67% of eyes within 0.50 D or less, respectively. IOL rotation averaged 0.25 ± 0.65 degrees, all eyes having rotation of less than 5 degrees. Contrast sensitivity was within or above normal levels under photopic and mesopic conditions, with or without glare, except for 12 cpd under mesopic conditions with glare. Questionnaire responses indicated 96.67% of patients were satisfied or very satisfied with postoperative vision, and 80.00% to 96.67% reported no difficulty in different daily activities.

CONCLUSIONS:

The Asqelio Trifocal Toric IOL demonstrated favorable outcomes, providing excellent visual performance at all distances, precise refractive results, and remarkable rotational stability. Patients reported high satisfaction levels and minimal difficulty in daily activities. [J Refract Surg. 2024;40(6)e407-e419.].
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Diseño de Prótesis / Refracción Ocular / Sensibilidad de Contraste / Visión Binocular / Agudeza Visual / Satisfacción del Paciente / Facoemulsificación / Seudofaquia / Implantación de Lentes Intraoculares / Lentes Intraoculares Multifocales Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Refract Surg Asunto de la revista: OFTALMOLOGIA Año: 2024 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Diseño de Prótesis / Refracción Ocular / Sensibilidad de Contraste / Visión Binocular / Agudeza Visual / Satisfacción del Paciente / Facoemulsificación / Seudofaquia / Implantación de Lentes Intraoculares / Lentes Intraoculares Multifocales Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Refract Surg Asunto de la revista: OFTALMOLOGIA Año: 2024 Tipo del documento: Article