Accuracy of Volumetric Bone Mineral Density Measurement in Weight Bearing, Cone Beam Computed Tomography.
Waungana, Tadiwa H; Qiu, Keven; Tse, Justin J; Anderson, Donald D; Emery, Carolyn A; Boyd, Steven K; Manske, Sarah L.
Afiliación
  • Waungana TH; Biomedical Engineering Graduate Program, University of Calgary, Alberta, Canada; McCaig Institute for Bone and Joint Health, University of Calgary, Alberta, Canada; Department of Radiology, Cumming School of Medicine, University of Calgary, Alberta, Canada.
  • Qiu K; McCaig Institute for Bone and Joint Health, University of Calgary, Alberta, Canada; David R. Cheriton School of Computer Science, University of Waterloo, Waterloo, Ontario, Canada.
  • Tse JJ; McCaig Institute for Bone and Joint Health, University of Calgary, Alberta, Canada; Department of Radiology, Cumming School of Medicine, University of Calgary, Alberta, Canada.
  • Anderson DD; Department of Orthopedics and Rehabilitation, University of Iowa, Iowa City, Iowa, United States.
  • Emery CA; McCaig Institute for Bone and Joint Health, University of Calgary, Alberta, Canada; Faculty of Kinesiology, University of Calgary, Alberta, Canada; Alberta Children's Hospital Research Institute, University of Calgary, Alberta, Canada; O'Brien Institute for Public Health, University of Calgary, Albe
  • Boyd SK; Biomedical Engineering Graduate Program, University of Calgary, Alberta, Canada; McCaig Institute for Bone and Joint Health, University of Calgary, Alberta, Canada; Department of Radiology, Cumming School of Medicine, University of Calgary, Alberta, Canada.
  • Manske SL; Biomedical Engineering Graduate Program, University of Calgary, Alberta, Canada; McCaig Institute for Bone and Joint Health, University of Calgary, Alberta, Canada; Department of Radiology, Cumming School of Medicine, University of Calgary, Alberta, Canada. Electronic address: smanske@ucalgary.ca.
J Clin Densitom ; 27(3): 101504, 2024.
Article en En | MEDLINE | ID: mdl-38897133
ABSTRACT

BACKGROUND:

Weight bearing computed tomography (WBCT) utilizes cone beam CT technology to provide assessments of lower limb joint structures while they are functionally loaded. Grey-scale values indicative of X-ray attenuation that are output from cone beam CT are challenging to calibrate, and their use for bone mineral density (BMD) measurement remains debatable. To determine whether WBCT can be reliably used for cortical and trabecular BMD assessment, we sought to establish the accuracy of BMD measurements at the knee using modern WBCT by comparing them to measurements from conventional CT.

METHODS:

A hydroxyapatite phantom with three inserts of varying densities was used to systematically quantify signal uniformity and BMD accuracy across the acquisition volume. We evaluated BMD in vivo (n = 5, female) using synchronous and asynchronous calibration techniques in WBCT and CT. To account for variation in attenuation along the height (z-axis) of acquisition volumes, we tested a height-dependent calibration approach for both WBCT and CT images.

RESULTS:

Phantom BMD measurement error in WBCT was as high as 15.3% and consistently larger than CT (up to 5.6%). Phantom BMD measures made under synchronous conditions in WBCT improved measurement accuracy by up to 3% but introduced more variability in measured BMD. We found strong correlations (R = 0.96) as well as wide limits of agreement (-324 mgHA/cm3 to 183 mgHA/cm3) from Bland-Altman analysis between WBCT and CT measures in vivo that were not improved by height-dependent calibration.

CONCLUSION:

Whilst BMD accuracy from WBCT was found to be dependent on apparent density, accuracy was independent of the calibration technique (synchronous or asynchronous) and the location of the measurement site within the field of view. Overall, we found strong correlations between BMD measures from WBCT and CT and in vivo measures to be more accurate in trabecular bone regions. Importantly, WBCT can be used to distinguish between anatomically relevant differences in BMD, however future work is necessary to determine the repeatability and sensitivity of BMD measures in WBCT.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Densidad Ósea / Soporte de Peso / Fantasmas de Imagen / Tomografía Computarizada de Haz Cónico Límite: Aged / Female / Humans / Middle aged Idioma: En Revista: J Clin Densitom Asunto de la revista: ORTOPEDIA Año: 2024 Tipo del documento: Article País de afiliación: Canadá

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Densidad Ósea / Soporte de Peso / Fantasmas de Imagen / Tomografía Computarizada de Haz Cónico Límite: Aged / Female / Humans / Middle aged Idioma: En Revista: J Clin Densitom Asunto de la revista: ORTOPEDIA Año: 2024 Tipo del documento: Article País de afiliación: Canadá