Your browser doesn't support javascript.
loading
Validation and clinical application of implantable loop recorders for diagnosis of atrial fibrillation in horses.
Kjeldsen, Sofie Troest; Nissen, Sarah D; Christensen, Nina C; Haugaard, Simon L; Schneider, Mélodie J; Vinther, Zenta; Sattler, Stefan M; Carstensen, Helena; Jøns, Christian; Hopster-Iversen, Charlotte; Buhl, Rikke.
Afiliación
  • Kjeldsen ST; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Nissen SD; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Christensen NC; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Haugaard SL; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Schneider MJ; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Vinther Z; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Sattler SM; Department of Cardiology, Herlev and Gentofte University Hospital, Gentofte, Denmark.
  • Carstensen H; Department of Biomedical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Jøns C; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
  • Hopster-Iversen C; Department of Cardiology, The Heart Centre, Copenhagen University Hospital, Rigshospitalet, Copenhagen, Denmark.
  • Buhl R; Department of Veterinary Clinical Sciences, Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark.
Equine Vet J ; 2024 Jun 21.
Article en En | MEDLINE | ID: mdl-39031582
ABSTRACT

BACKGROUND:

Paroxysmal atrial fibrillation (pAF) occurs sporadically and can impair athletic performance. Gold standard for diagnosis is surface electrocardiography (ECG), however, this requires AF to be sustained. Implantable loop recorders (ILRs) are routinely used for AF detection in human medicine. While ILR placement has been studied in horses, its AF detection performance is unknown.

OBJECTIVES:

(I) Validation of ILRs for AF detection in horses. (II) Determining pAF incidence using ILRs and estimate the positive predictive value (PPV). STUDY

DESIGN:

(I) Experimental study; (II) Longitudinal observational study.

METHODS:

(I) Implantation of ILRs in 15 horses with AF and 13 horses in sinus rhythm. Holter ECGs were recorded at 1, 4, 8, 12 and 16 weeks of AF. The ILR ECGs were compared with surface ECGs to assess diagnostic sensitivity and specificity. (II) Eighty horses (43 Warmbloods, 37 Standardbreds) with ILRs were monitored for 367 days [IQR 208-621].

RESULTS:

(I) ILRs detected AF on all recording days, in horses with AF, with a sensitivity of 66.1% (95% CI 65.8-66.5) and a specificity of 99.99% (95% CI 99.97-99.99). The sensitivity remained consistent across all time points. (II) The incidence of pAF was 6.3% (5/80). In horses with pAF, the PPV ranged from 8% to 87%. Increased body condition score (BCS > 6/9) was associated with an increased number of false positive episodes (p = 0.005). MAIN

LIMITATIONS:

(I) Horses were stabled during the ECG recordings, and AF was induced, rather than naturally occurring pAF. (II) Integrated algorithm in this ILR is optimised for AF detection in humans using remote monitors. Additionally, sensing is affected by motion artefacts.

CONCLUSION:

The ILR reliably detected AF in resting horses, particularly in horses with normal BCS (6/9). The ILR proved useful to detect pAF and is recommended alongside Holter monitoring for diagnostic workup of horses with suspected pAF.