Performance of iodine quantification through high-pitch dual-source photon-counting CT: a phantom study.
Liu, Peng; Zhou, Shanshui; Dong, Haipeng; Li, Jiqiang; Xu, Zhihan; Lin, Shushen; Yang, Wenjie; Yan, Fuhua; Qin, Le.
Afiliación
  • Liu P; Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China.
  • Zhou S; Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China.
  • Dong H; Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China.
  • Li J; Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China.
  • Xu Z; CT Collaboration, Siemens Healthineers, 399 West Haiyang Road, Shanghai, 200126, China.
  • Lin S; CT Collaboration, Siemens Healthineers, 399 West Haiyang Road, Shanghai, 200126, China.
  • Yang W; Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China.
  • Yan F; Department of Radiology, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, No. 197 Ruijin Er Road, Shanghai, 200025, China. yfh11655@rjh.com.cn.
  • Qin L; Faculty of Medical Imaging Technology, College of Health Science and Technology, Shanghai Jiao Tong University School of Medicine, Shanghai, China. yfh11655@rjh.com.cn.
Jpn J Radiol ; 2024 Oct 09.
Article en En | MEDLINE | ID: mdl-39382795
ABSTRACT

PURPOSE:

To investigate the feasibility and accuracy of iodine quantification using PCD-CT in standard-pitch and high-pitch scanning at different scan parameters in a phantom model. MATERIALS AND

METHODS:

Four inserts with known iodine concentrations (2, 5, 10, and 15 mg/mL) were placed in the removable CT phantom and scanned using high-pitch (3.2) and standard-pitch (0.8) modes on PCD-CT. Two tube voltages (120 and 140 kVp) and four radiation doses (1, 3, 5, and 10 mGy) were alternated. Each scan setting was repeated three times. Mean iodine concentration for each insert across three consecutive slices was recorded. Percentage absolute bias (PAB) was assessed for iodine quantification.

RESULTS:

A total of 96 acquisitions were conducted. In small phantom, the average for PAB was 2.96% (range 1.75% ~ 4.56%) and 1.67% (range 1.00% ~ 3.42%) for high-pitch and standard-pitch acquisitions, respectively. In large phantom, it was 3.72% (range 1.75% ~ 5.97%) and 2.94% (range 1.75% ~ 4.70%). Linear regression analysis revealed that only phantom size significantly influenced (P < 0.001) the accuracy of iodine quantification.

CONCLUSION:

The high-pitch scan mode in PCD-CT can be used to quantify iodine density with similar accuracy compared with standard pitch.