1.
IEEE Access ; 9: 25946-25958, 2021.
Artigo em Inglês | MEDLINE | ID: mdl-33996341

RESUMO

Proton CT (pCT) is a promising new imaging technique that can reconstruct relative stopping power (RSP) more accurately than x-ray CT in each cubic millimeter voxel of the patient. This, in turn, will result in better proton range accuracy and, therefore, smaller planned tumor volumes (PTV). The hardware description and some reconstructed images have previously been reported. In a series of two contributions, we focus on presenting the software algorithms that convert pCT detector data to the final reconstructed pCT images for application in proton treatment planning. There were several options on how to accomplish this, and we will describe our solutions at each stage of the data processing chain. In the first paper of this series, we present the data acquisition with the pCT tracking and energy-range detectors and how the data are preprocessed, including the conversion to the well-formatted track information from tracking data and water-equivalent path length from the data of a calibrated multi-stage energy-range detector. These preprocessed data are then used for the initial image formation with an FDK cone-beam CT algorithm. The output of data acquisition, preprocessing, and FDK reconstruction is presented along with illustrative imaging results for two phantoms, including a pediatric head phantom. The second paper in this series will demonstrate the use of iterative solvers in conjunction with the superiorization methodology to further improve the images resulting from the upfront FDK image reconstruction and the implementation of these algorithms on a hybrid CPU/GPU computer cluster.

2.
Phys Med Biol ; 64(15): 15NT01, 2019 08 01.
Artigo em Inglês | MEDLINE | ID: mdl-31158820

RESUMO

Data filtering is crucial for accurate relative stopping power (RSP) reconstruction in proton CT (pCT). In this work, we assess different filters and their performance for the US pCT collaboration prototype pCT system in Monte Carlo (MC) simulations. The potential of using the recently proposed [Formula: see text]-E filter for removing nuclear interactions that occurred in the energy/range detector of the pCT system is investigated. Full pCT scans were acquired with the TOPAS MC simulated version of the prototype scanner that comprises two tracking detectors and a five stage energy/range detector. An ideal water cylinder and a water cylinder with five tissue inserts were investigated. Before image reconstruction, a [Formula: see text] WEPL filter was applied as the only filter, or in addition to filters acting on the energy deposit in each of the energy detector stages, as done currently with the prototype. The potential of the [Formula: see text]-E filter that was recently proposed for helium imaging was assessed. The results were compared to simulations for which nuclear interactions were disabled representing ground truth. The [Formula: see text] WEPL filter alone was not sufficient to filter out all nuclear interaction events and systematic fluctuations in the form of ring artifacts were present in the pCT reconstructed images. Applying energy filters currently used with the device prior to the [Formula: see text] WEPL filter only slightly improved the image quality. A [Formula: see text] WEPL filter improved the mean RSP accuracy, but could not fully remove the systematic fluctuations. The [Formula: see text]-E filter in addition to the current reconstruction procedure efficiently removed the systematic fluctuations and the achieved RSP accuracy closely matched the simulation without nuclear interactions. This study demonstrates the dependence of the accuracy of the usual [Formula: see text] WEPL filter on uncertainties arising within the energy detector. By enabling to remove such uncertainties, the [Formula: see text]-E method proved to yield some potential for improving the accuracy of pCT.


Assuntos
Prótons , Tomografia/instrumentação , Método de Monte Carlo , Tomografia/métodos
3.
Z Med Phys ; 18(4): 286-96, 2008.
Artigo em Inglês | MEDLINE | ID: mdl-19205298

RESUMO

Evaluation and monitoring of the cancer risk from space radiation exposure is a crucial requirement for the success of long-term space missions. One important task in the risk calculation is to properly weigh the various components of space radiation dose according to their assumed contribution to the cancer risk relative to the risk associated with radiation of low ionization density. Currently, quality factors of radiation both on the ground and in space are defined by national and international commissions based on existing radiobiological data and presumed knowledge of the ionization density distribution of the radiation field at a given point of interest. This approach makes the determination of the average quality factor ofa given radiation field a rather complex task. In this contribution, we investigate the possibility to define quality factors of space radiation exposure based on nanodosimetric data. The underlying formalism of the determination of quality factors on the basis of nanodosimetric data is described, and quality factors for protons and ions (helium and carbon) of different energies based on simulated nanodosimetric data are presented. The value and limitations of this approach are discussed.


Assuntos
Dano ao DNA , Monitoramento de Radiação/normas , Proteção Radiológica/normas , Radiometria/normas , Voo Espacial , Humanos , Transferência Linear de Energia/efeitos da radiação , Probabilidade , Doses de Radiação , Radiometria/efeitos adversos , Medição de Risco , Sensibilidade e Especificidade
4.
Phys Med Biol ; 63(19): 195016, 2018 10 02.
Artigo em Inglês | MEDLINE | ID: mdl-30183679

RESUMO

Single-event ion imaging enables the direct reconstruction of the relative stopping power (RSP) information required for ion-beam therapy. Helium ions were recently hypothesized to be the optimal species for such technique. The purpose of this work is to investigate the effect of secondary fragments on the image quality of helium CT (HeCT) and to assess the performance of a prototype proton CT (pCT) scanner when operated with helium beams in Monte Carlo simulations and experiment. Experiments were conducted installing the U.S. pCT consortium prototype scanner at the Heidelberg Ion-Beam Therapy Center (HIT). Simulations were performed with the scanner using the TOPAS toolkit. HeCT images were reconstructed for a cylindrical water phantom, the CTP404 (sensitometry), and the CTP528 (line-pair) [Formula: see text] ® modules. To identify and remove individual events caused by fragmentation, the multistage energy detector of the scanner was adapted to function as a [Formula: see text] telescope. The use of the developed filter eliminated the otherwise arising ring artifacts in the HeCT reconstructed images. For the HeCT reconstructed images of a water phantom, the maximum RSP error was improved by almost a factor 8 with respect to unfiltered images in the simulation and a factor 10 in the experiment. Similarly, for the CTP404 module, the mean RSP accuracy improved by a factor 6 in both the simulation and the experiment when the filter was applied (mean relative error 0.40% in simulation, 0.45% in experiment). In the evaluation of the spatial resolution through the CTP528 module, the main effect of the filter was noise reduction. For both simulated and experimental images the spatial resolution was ∼4 lp cm-1. In conclusion, the novel filter developed for secondary fragments proved to be effective in improving the visual quality and RSP accuracy of the reconstructed images. With the filter, the pCT scanner is capable of accurate HeCT imaging.


Assuntos
Hélio , Processamento de Imagem Assistida por Computador/métodos , Cintilografia/métodos , Humanos , Método de Monte Carlo , Imagens de Fantasmas , Cintilografia/normas
5.
Med Phys ; 44(3): 1002-1016, 2017 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-28094862

RESUMO

PURPOSE: Proton computed tomography (pCT) is a promising imaging technique to substitute or at least complement x-ray CT for more accurate proton therapy treatment planning as it allows calculating directly proton relative stopping power from proton energy loss measurements. A proton CT scanner with a silicon-based particle tracking system and a five-stage scintillating energy detector has been completed. In parallel a modular software platform was developed to characterize the performance of the proposed pCT. METHOD: The modular pCT software platform consists of (1) a Geant4-based simulation modeling the Loma Linda proton therapy beam line and the prototype proton CT scanner, (2) water equivalent path length (WEPL) calibration of the scintillating energy detector, and (3) image reconstruction algorithm for the reconstruction of the relative stopping power (RSP) of the scanned object. In this work, each component of the modular pCT software platform is described and validated with respect to experimental data and benchmarked against theoretical predictions. In particular, the RSP reconstruction was validated with both experimental scans, water column measurements, and theoretical calculations. RESULTS: The results show that the pCT software platform accurately reproduces the performance of the existing prototype pCT scanner with a RSP agreement between experimental and simulated values to better than 1.5%. CONCLUSIONS: The validated platform is a versatile tool for clinical proton CT performance and application studies in a virtual setting. The platform is flexible and can be modified to simulate not yet existing versions of pCT scanners and higher proton energies than those currently clinically available.


Assuntos
Simulação por Computador , Prótons , Software , Tomografia/instrumentação , Tomografia/métodos , Algoritmos , Calibragem , Criança , Desenho de Equipamento , Cabeça/diagnóstico por imagem , Humanos , Modelos Anatômicos , Modelos Teóricos , Terapia com Prótons/instrumentação , Terapia com Prótons/métodos , Tórax/diagnóstico por imagem , Água
6.
Nucl Instrum Methods Phys Res A ; 809: 120-129, 2016 Feb 11.
Artigo em Inglês | MEDLINE | ID: mdl-26957679

RESUMO

Radiation therapy with protons and heavier ions is an attractive form of cancer treatment that could enhance local control and survival of cancers that are currently difficult to cure and lead to less side effects due to sparing of normal tissues. However, particle therapy faces a significant technical challenge because one cannot accurately predict the particle range in the patient using data provided by existing imaging technologies. Proton computed tomography (pCT) is an emerging imaging modality capable of improving the accuracy of range prediction. In this paper, we describe the successive pCT scanners designed and built by our group with the goal to support particle therapy treatment planning and image guidance by reconstructing an accurate 3D map of the stopping power relative to water in patient tissues. The pCT scanners we have built to date consist of silicon telescopes, which track the proton before and after the object to be reconstructed, and an energy or range detector, which measures the residual energy and/or range of the protons used to evaluate the water equivalent path length (WEPL) of each proton in the object. An overview of a decade-long evolution of the conceptual design of pCT scanners and their calibration is given. Results of scanner performance tests are presented, which demonstrate that the latest pCT scanner approaches readiness for clinical applications in hadron therapy.