1.
Magn Reson Med ; 90(6): 2627-2642, 2023 12.
Artigo em Inglês | MEDLINE | ID: mdl-37533196

RESUMO

PURPOSE: The purpose of this study is to present a strategy to calculate the implant-friendly (IF) excitation modes-which mitigate the RF heating at the contacts of deep brain stimulation (DBS) electrodes-of multichannel RF coils at 7 T. METHODS: An induced RF current on an implantable electrode generates a scattered magnetic field whose left-handed circularly polarizing component ( B 1 + $$ B{1}^{+} $$ ) is approximated using a B 1 + $$ B{1}^{+} $$ -mapping technique and subsequently used as a gauge for the electrode's induced current. Using this approach, the relative induced currents resulting from each channel of a multichannel RF coil on the DBS electrode were calculated. The IF modes of the corresponding multichannel coil were determined by calculating the null space of the relative induced currents. The proposed strategy was tested and validated for unilateral and bilateral commercial DBS electrodes (directional lead; Infinity DBS system, Abbott Laboratories) placed inside a uniform phantom by performing heating and imaging studies on a 7T MRI scanner using a 16-channel transceive RF coil. RESULTS: Neither individual IF modes nor shim solutions obtained from IF modes induced significant temperature increase when used for a high-power turbo spin-echo sequence. In contrast, shimming with the scanner's toolbox (i.e., based on per-channel B 1 + $$ B{1}^{+} $$ fields) resulted in a more than 2°C temperature increase for the same amount of input power. CONCLUSION: A strategy for calculating the IF modes of a multichannel RF coil is presented. This strategy was validated using a 16-channel RF coil at 7 T for unilateral and bilateral commercial DBS electrodes inside a uniform phantom.


Assuntos
Estimulação Encefálica Profunda , Estimulação Encefálica Profunda/métodos , Imageamento por Ressonância Magnética/métodos , Eletrodos Implantados , Imagens de Fantasmas , Ondas de Rádio
2.
Magn Reson Med ; 89(6): 2204-2216, 2023 06.
Artigo em Inglês | MEDLINE | ID: mdl-36669882

RESUMO

PURPOSE: To expand on the previously developed B 1 + $$ {\mathrm{B}}_1^{+} $$ -encoding technique, frequency-modulated Rabi-encoded echoes (FREE), to perform accelerated image acquisition by collecting multiple lines of k-space in an echo train. METHODS: FREE uses adiabatic full-passage pulses and a spatially varying RF field to encode unique spatial information without the use of traditional B0 gradients. The original implementation relied on acquiring single lines of k-space, leading to long acquisitions. In this work, an acceleration scheme is presented in which multiple echoes are acquired in a single shot, analogous to conventional fast spin-echo sequences. Theoretical analysis and computer simulations investigated the feasibility of this approach and presented a framework to analyze important imaging parameters of FREE-based sequences. Experimentally, the multi-echo approach was compared with conventional phase-encoded images of the human visual cortex using a simple surface transceiver coil. Finally, different contrasts demonstrated the clinical versatility of the new accelerated sequence. RESULTS: Images were acquired with an acceleration factor of 3.9, compared with the previous implementation of FREE, without exceeding specific absorption rate limits. Different contrasts can easily be acquired without major modifications, including inversion recovery-type images. CONCLUSION: FREE initially illustrated the feasibility of performing slice-selective 2D imaging of the human brain without the need for a B0 gradient along the y-direction. The multi-echo version maintains the advantages that B 1 + $$ {\mathrm{B}}_1^{+} $$ encoding provides but represents an important step toward improving the clinical feasibility of such sequences. Additional acceleration and more advanced reconstruction techniques could further improve the clinical viability of FREE-based techniques.


Assuntos
Encéfalo , Imageamento por Ressonância Magnética , Humanos , Imageamento por Ressonância Magnética/métodos , Encéfalo/diagnóstico por imagem , Cabeça , Simulação por Computador , Aceleração , Processamento de Imagem Assistida por Computador/métodos , Imagens de Fantasmas
3.
Magn Reson Med ; 89(5): 2100-2108, 2023 05.
Artigo em Inglês | MEDLINE | ID: mdl-36517956

RESUMO

PURPOSE: To correct image distortions that result from nonlinear spatial variation in the transmit RF field amplitude ( B 1 + $$ {B}_1^{+} $$ ) when performing spatial encoding with the method called frequency-modulated Rabi encoded echoes (FREE). THEORY AND METHODS: An algorithm developed to correct image distortion resulting from the use of nonlinear static field (B0 ) gradients in standard MRI is adapted herein to correct image distortion arising from a nonlinear B 1 + $$ {B}_1^{+} $$ -gradient field in FREE. From a B 1 + $$ {B}_1^{+} $$ -map, the algorithm performs linear interpolation and intensity scaling to correct the image. The quality of the distortion correction is evaluated in 1.5T images of a grid phantom and human occipital lobe. RESULTS: An expanded theoretical description of FREE revealed the symmetry between this B 1 + $$ {B}_1^{+} $$ -gradient field spatial-encoding and standard B0 -gradient field spatial-encoding. The adapted distortion-correction algorithm substantially reduced image distortions arising in the spatial dimension that was encoded by the nonlinear B 1 + $$ {B}_1^{+} $$ gradient of a circular surface coil. CONCLUSION: Image processing based on straightforward linear interpolation and intensity scaling, as previously applied in conventional MRI, can effectively reduce distortions in FREE images acquired with nonlinear B 1 + $$ {B}_1^{+} $$ -gradient fields.


Assuntos
Processamento de Imagem Assistida por Computador , Imageamento por Ressonância Magnética , Humanos , Imageamento por Ressonância Magnética/métodos , Processamento de Imagem Assistida por Computador/métodos , Algoritmos , Imagens de Fantasmas
4.
Magn Reson Med ; 87(2): 674-685, 2022 02.
Artigo em Inglês | MEDLINE | ID: mdl-34498768

RESUMO

PURPOSE: Reduce expense and increase accessibility of MRI by eliminating pulsed field (B0 ) gradient hardware. METHODS: A radiofrequency imaging method is described that enables spatial encoding without B0 gradients. This method, herein referred to as frequency-modulated Rabi-encoded echoes (FREE), utilizes adiabatic full passage pulses and a gradient in the RF field (B1 ) to produce spatially dependent phase modulation, equivalent to conventional phase encoding. In this work, Cartesian phase encoding was accomplished using FREE in a multi-shot double spin-echo sequence. Theoretical analysis and computer simulations investigated the influence of resonance offset and B1 -gradient steepness and magnitude on reconstruction quality, which limit other radiofrequency imaging methodologies. Experimentally, FREE was compared to conventional phase-encoded MRI on human visual cortex using a simple surface transceiver coil. RESULTS: Image distortions occurred in FREE when using nonlinear B1 fields where the phase dependence becomes nonlinear, but with minimal change in signal intensity. Resonance offset effects were minimal for Larmor frequencies within the adiabatic full-passage pulse bandwidth. CONCLUSION: For the first time, FREE enabled slice-selective 2D imaging of the human brain without a B0 gradient in the y-direction. FREE achieved high resolution in regions where the B1 gradient was steepest, whereas images were distorted in regions where nonlinearity in the B1 gradient was significant. Given that FREE experiences no significant signal loss due to B1 nonlinearities and resonance offset, image distortions shown in this work might be corrected in the future based on B1 and B0 maps.


Assuntos
Imageamento por Ressonância Magnética , Ondas de Rádio , Encéfalo/diagnóstico por imagem , Simulação por Computador , Humanos , Imagens de Fantasmas
5.
Magn Reson Med ; 87(4): 2074-2088, 2022 04.
Artigo em Inglês | MEDLINE | ID: mdl-34825735

RESUMO

PURPOSE: The purpose of this study is to introduce a new antenna element with improved transmit performance, named the nonuniform dielectric substrate (NODES) antenna, for building transmit arrays at ultrahigh-field. METHODS: We optimized a dipole antenna at 10.5 Tesla by maximizing the B1+ -SAR efficiency in a phantom for a human spine target. The optimization parameters included permittivity variation in the substrate, substrate thickness, antenna length, and conductor geometry. We conducted electromagnetic simulations as well as phantom experiments to compare the transmit/receive performance of the proposed NODES antenna design with existing coil elements from the literature. RESULTS: Single NODES element showed up to 18% and 30% higher B1+ -SAR efficiency than the fractionated dipole and loop elements, respectively. The new element is substantially shorter than a commonly used dipole, which enables z-stacked array formation; it is additionally capable of providing a relatively uniform current distribution along its conductors. The nine-channel transmit/receive NODES array achieved 7.5% higher B1+ homogeneity than a loop array with the same number of elements. Excitation with the NODES array resulted in 33% lower peak 10g-averaged SAR and required 34% lower input power than the loop array for the target anatomy of the spine. CONCLUSION: In this study, we introduced a new RF coil element: the NODES antenna. NODES antenna outperformed the widely used loop and dipole elements and may provide improved transmit/receive performance for future ultrahigh field MRI applications.


Assuntos
Imageamento por Ressonância Magnética , Ondas de Rádio , Desenho de Equipamento , Humanos , Imagens de Fantasmas , Coluna Vertebral/diagnóstico por imagem
6.
IEEE Antennas Wirel Propag Lett ; 21(9): 1857-1861, 2022 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-37020750

RESUMO

In this letter, we evaluate antenna designs for ultra-high frequency and field (UHF) human brain magnetic resonance imaging (MRI) at 10.5 tesla (T). Although MRI at such UHF is expected to provide major signal-to-noise gains, the frequency of interest, 447 MHz, presents us with challenges regarding improved B1 + efficiency, image homogeneity, specific absorption rate (SAR), and antenna element decoupling for array configurations. To address these challenges, we propose the use of both monopole and dipole antennas in a novel hybrid configuration, which we refer to as a mono-dipole hybrid antenna (MDH) array. Compared to an 8-channel dipole antenna array of the same dimensions, the 8-channel MDH array showed an improvement in decoupling between adjacent array channels, as well as ~18% higher B1 + and SAR efficiency near the central region of the phantom based on simulation and experiment. However, the performances of the MDH and dipole antenna arrays were overall similar when evaluating a human model in terms of peak B1 + efficiency, 10 g SAR, and SAR efficiency. Finally, the concept of an MDH array showed an advantage in improved decoupling, SAR, and B1 + near the superior region of the brain for human brain imaging.

7.
Magn Reson Med ; 85(6): 3522-3530, 2021 06.
Artigo em Inglês | MEDLINE | ID: mdl-33464649

RESUMO

PURPOSE: In this work, we investigated how the position of the radiofrequency (RF) shield can affect the signal-to-noise ratio (SNR) of a receive RF coil. Our aim was to obtain physical insight for the design of a 10.5T 32-channel head coil, subject to the constraints on the diameter of the RF shield imposed by the head gradient coil geometry. METHOD: We used full-wave numerical simulations to investigate how the SNR of an RF receive coil depends on the diameter of the RF shield at ultra-high magnetic field (UHF) strengths (≥7T). RESULTS: Our simulations showed that there is an SNR-optimal RF shield size at UHF strength, whereas at low field the SNR monotonically increases with the shield diameter. For a 32-channel head coil at 10.5T, an optimally sized RF shield could act as a cylindrical waveguide and increase the SNR in the brain by 27% compared to moving the shield as far as possible from the coil. Our results also showed that a separate transmit array between the RF shield and the receive array could considerably reduce SNR even if they are decoupled. CONCLUSION: At sufficiently high magnetic field strength, the design of local RF coils should be optimized together with the design of the RF shield to benefit from both near field and resonant modes.


Assuntos
Imageamento por Ressonância Magnética , Ondas de Rádio , Encéfalo/diagnóstico por imagem , Desenho de Equipamento , Cabeça , Imagens de Fantasmas , Razão Sinal-Ruído
8.
Magn Reson Med ; 86(6): 3292-3303, 2021 12.
Artigo em Inglês | MEDLINE | ID: mdl-34272898

RESUMO

PURPOSE: Investigating the designs and effects of high dielectric constant (HDC) materials in the shape of a conformal helmet on the enhancement of RF field and reduction of specific absorption rate at 10.5 T for human brain studies. METHODS: A continuous and a segmented four-piece HDC helmet fit to a human head inside an eight-channel fractionated-dipole array were constructed and studied with a phantom and a human head model using computer electromagnetic simulations. The simulated transmit efficiency and receive sensitivity were experimentally validated using a phantom with identical electric properties and helmet-coil configurations of the computer model. The temporal and spatial distributions of displacement currents on the HDC helmets were analyzed. RESULTS: Using the continuous HDC helmet, simulation results in the human head model demonstrated an average transmit efficiency enhancement of 66%. A propagating displacement current was induced on the continuous helmet, leading to an inhomogeneous RF field enhancement in the brain. Using the segmented four-piece helmet design to reduce this effect, an average 55% and 57% enhancement in the transmit efficiency and SNR was achieved in human head, respectively, along with 8% and 28% reductions in average and maximum local specific absorption rate. CONCLUSION: The HDC helmets enhanced the transmit efficiency and SNR of the dipole array coil in the human head at 10.5 T. The segmentation of the helmet to disrupt the continuity of circumscribing displacement currents in the helmet produced a more uniform distribution of the transmit field and lower specific absorption rate in the human head compared with the continuous helmet design.


Assuntos
Dispositivos de Proteção da Cabeça , Imageamento por Ressonância Magnética , Encéfalo/diagnóstico por imagem , Desenho de Equipamento , Humanos , Imagens de Fantasmas , Ondas de Rádio
9.
NMR Biomed ; 34(4): e4472, 2021 04.
Artigo em Inglês | MEDLINE | ID: mdl-33511726

RESUMO

A 32-channel RF coil was developed for brain imaging of anesthetized non-human primates (rhesus macaque) at 10.5 T. The coil is composed of an 8-channel dipole transmit/receive array, close-fitting 16-channel loop receive array headcap, and 8-channel loop receive array lower insert. The transceiver dipole array is composed of eight end-loaded dipole elements self-resonant at the 10.5 T proton Larmor frequency. These dipole elements were arranged on a plastic cylindrical former, which was split into two to allow for convenient animal positioning. Nested into the bottom of the dipole array former is located an 8-channel loop receive array, which contains 5 × 10 cm2 square loops arranged in two rows of four loops. Arranged in a close-fitting plastic headcap is located a high-density 16-channel loop receive array. This array is composed of 14 round loops 37 mm in diameter and 2 partially detachable, irregularly shaped loops that encircle the ears. Imaging experiments were performed on anesthetized non-human primates on a 10.5 T MRI system equipped with body gradients with a 60 cm open bore. The coil enabled submillimeter (0.58 mm isotropic) high-resolution anatomical and functional imaging as well as tractography of fasciculated axonal bundles. The combination of a close-fitting loop receive array and dipole transceiver array allowed for a higher-channel-count receiver and consequent higher signal-to-noise ratio and parallel imaging gains. Parallel imaging performance supports high-resolution functional MRI and diffusion MRI with a factor of three reduction in sampling. The transceive array elements during reception contributed approximately one-quarter of the signal-to-noise ratio in the lower half of the brain, which was farthest from the close-fitting headcap receive array.


Assuntos
Cabeça/diagnóstico por imagem , Imageamento por Ressonância Magnética/métodos , Animais , Feminino , Macaca mulatta , Razão Sinal-Ruído
10.
Sensors (Basel) ; 21(21)2021 Oct 30.
Artigo em Inglês | MEDLINE | ID: mdl-34770558

RESUMO

For ultra-high field and frequency (UHF) magnetic resonance imaging (MRI), the associated short wavelengths in biological tissues leads to penetration and homogeneity issues at 10.5 tesla (T) and require antenna transmit arrays for efficiently generated 447 MHz B1+ fields (defined as the transmit radiofrequency (RF) magnetic field generated by RF coils). Previously, we evaluated a 16-channel combined loop + dipole antenna (LD) 10.5 T head array. While the LD array configuration did not achieve the desired B1+ efficiency, it showed an improvement of the specific absorption rate (SAR) efficiency compared to the separate 8-channel loop and separate 8-channel dipole antenna arrays at 10.5 T. Here we compare a 16-channel dipole antenna array with a 16-channel LD array of the same dimensions to evaluate B1+ efficiency, 10 g SAR, and SAR efficiency. The 16-channel dipole antenna array achieved a 24% increase in B1+ efficiency in the electromagnetic simulation and MR experiment compared to the LD array, as measured in the central region of a phantom. Based on the simulation results with a human model, we estimate that a 16-channel dipole antenna array for human brain imaging can increase B1+ efficiency by 15% with similar SAR efficiency compared to a 16-channel LD head array.


Assuntos
Imageamento por Ressonância Magnética , Ondas de Rádio , Encéfalo/diagnóstico por imagem , Desenho de Equipamento , Humanos , Imagens de Fantasmas
11.
Sensors (Basel) ; 21(18)2021 Sep 08.
Artigo em Inglês | MEDLINE | ID: mdl-34577210

RESUMO

For human head magnetic resonance imaging at 10.5 tesla (T), we built an 8-channel transceiver dipole antenna array and evaluated the influence of coaxial feed cables. The influence of coaxial feed cables was evaluated in simulation and compared against a physically constructed array in terms of transmit magnetic field (B1+) and specific absorption rate (SAR) efficiency. A substantial drop (23.1% in simulation and 20.7% in experiment) in B1+ efficiency was observed with a tight coaxial feed cable setup. For the investigation of the feed location, the center-fed dipole antenna array was compared to two 8-channel end-fed arrays: monopole and sleeve antenna arrays. The simulation results with a phantom indicate that these arrays achieved ~24% higher SAR efficiency compared to the dipole antenna array. For a human head model, we observed 30.8% lower SAR efficiency with the 8-channel monopole antenna array compared to the phantom. Importantly, our simulation with the human model indicates that the sleeve antenna arrays can achieve 23.8% and 21% higher SAR efficiency compared to the dipole and monopole antenna arrays, respectively. Finally, we obtained high-resolution human cadaver images at 10.5 T with the 8-channel sleeve antenna array.


Assuntos
Cabeça , Imageamento por Ressonância Magnética , Simulação por Computador , Desenho de Equipamento , Cabeça/diagnóstico por imagem , Humanos , Imagens de Fantasmas
12.
Magn Reson Med ; 84(4): 1747-1762, 2020 10.
Artigo em Inglês | MEDLINE | ID: mdl-32115756

RESUMO

PURPOSE: In this study, we sought to develop a self-navigation strategy for improving the reconstruction of diffusion weighted 3D multishot echo planar imaging (EPI). We propose a method for extracting the phase correction information from the acquisition itself, eliminating the need for a 2D navigator, further accelerating the acquisition. METHODS: In-vivo acquisitions at 3T with 0.9 mm and 1.5 mm isotropic resolutions were used to evaluate the performance of the self-navigation strategy. Sensitivity to motion was tested using a large difference in pitch position of the head. Using a multishell diffusion weighted acquisition, tractography results were obtained at (0.9 mm)3 to validate the quality with conventional acquisition. RESULTS: The use of 3D multislab EPI with self-navigation enables 3D diffusion-weighted spin echo EPI acquisitions that have the same efficiency as 2D single-shot acquisition. For matched acquisition time the image signal-to-noise ratio (SNR) between 3D and 2D acquisition is shown to be comparable for whole-brain coverage with (1.5 mm)3 resolution and for (0.9 mm)3 resolution the 3D acquisition has higher SNR than what can be obtained with 2D acquisitions using current state-of-art multiband techniques. The self-navigation technique was shown to be stable under inter-volume motion. In tractography analysis, the higher resolution afforded by our technique enabled clear delineation of the tapetum and posterior corona radiata. CONCLUSION: The proposed self-navigation approach utilized a self-consistent phase in 3D diffusion weighted acquisitions. Its efficiency and stability were demonstrated for a plurality of common acquisitions. The proposed self-navigation approach allows for faster acquisition of 3D multishot EPI desirable for large field of view and/or higher resolution.


Assuntos
Imagem Ecoplanar , Interpretação de Imagem Assistida por Computador , Algoritmos , Encéfalo/diagnóstico por imagem , Imagem de Difusão por Ressonância Magnética , Reprodutibilidade dos Testes
13.
Magn Reson Med ; 84(6): 3485-3493, 2020 12.
Artigo em Inglês | MEDLINE | ID: mdl-32767392

RESUMO

PURPOSE: In this study, we investigate a strategy to reduce the local specific absorption rate (SAR) while keeping B1+ constant inside the region of interest (ROI) at the ultra-high field (B0 ≥ 7T) MRI. METHODS: Locally raising the resonance structure under the discontinuity (i.e., creating a bump) increases the distance between the accumulated charges and the tissue. As a result, it reduces the electric field and local SAR generated by these charges inside the tissue. The B1+ at a point that is sufficiently far from the coil, however, is not affected by this modification. In this study, three different resonant elements (i.e., loop coil, snake antenna, and fractionated dipole [FD]) are investigated. For experimental validation, a bumped FD is further investigated at 10.5T. After the validation, the transmit performances of eight-channel arrays of each element are compared through electromagnetic (EM) simulations. RESULTS: Introducing a bump reduced the peak 10g-averaged SAR by 21, 26, 23% for the loop and snake antenna at 7T, and FD at 10.5T, respectively. In addition, eight-channel bumped FD array at 10.5T had a 27% lower peak 10g-averaged SAR in a realistic human body simulation (i.e., prostate imaging) compared to an eight-channel FD array. CONCLUSION: In this study, we investigated a simple design strategy based on adding bumps to a resonant element to reduce the local SAR while maintaining B1+ inside an ROI. As an example, we modified an FD and performed EM simulations and phantom experiments with a 10.5T scanner. Results show that the peak 10g-averaged SAR can be reduced more than 25%.


Assuntos
Imageamento por Ressonância Magnética , Ondas de Rádio , Desenho de Equipamento , Humanos , Masculino , Imagens de Fantasmas , Próstata
14.
Magn Reson Med ; 84(6): 2953-2963, 2020 12.
Artigo em Inglês | MEDLINE | ID: mdl-32544274

RESUMO

PURPOSE: Spatial encoding for MRI is generally based on linear x, y, and z magnetic field gradients generated by a set of dedicated gradient coils. We recently introduced the dynamic multicoil technique (DYNAMITE) for B0 field control and demonstrated DYNAMITE MRI in a preclinical MR environment. In this study, we report the first realization of DYNAMITE MRI of the in vivo human head. METHODS: Gradient fields for DYNAMITE MRI were generated with a 28-channel multicoil hardware arranged in 4 rows of 7 coils on a cylindrical surface (length 359 mm, diameter 344 mm, maximum 5 A per coil). DYNAMITE MRIs of a resolution phantom and in vivo human heads were acquired with multislice gradient-echo, multislice spin-echo, and 3D gradient-echo sequences. The resultant image fidelity was compared to that obtained with conventional gradient coil technology. RESULTS: DYNAMITE field control enabled the realization of all imaging sequences with average gradient errors ≤ 1%. DYNAMITE MRI provided image quality and sensitivity comparable to conventional gradient technology without any obvious artifacts. Some minor geometric deformations were noticed primarily in the image periphery as the result of regional field imperfections. The imperfections can be readily approximated theoretically through numerical integration of the Biot-Savart law and removed through image distortion correction. CONCLUSION: The first realization of DYNAMITE MRI of the in vivo human head has been presented. The obtained image fidelity is comparable to MRI with conventional gradient coils, paving the way for full-fledged DYNAMITE MRI and B0 shim systems for human applications.


Assuntos
Encéfalo , Imageamento por Ressonância Magnética , Artefatos , Encéfalo/diagnóstico por imagem , Humanos , Campos Magnéticos , Imagens de Fantasmas
15.
Magn Reson Med ; 84(1): 484-496, 2020 07.
Artigo em Inglês | MEDLINE | ID: mdl-31751499

RESUMO

PURPOSE: The purpose of this study is to safely acquire the first human head images at 10.5T. METHODS: To ensure safety of subjects, we validated the electromagnetic simulation model of our coil. We obtained quantitative agreement between simulated and experimental B1+ and specific absorption rate (SAR). Using the validated coil model, we calculated radiofrequency power levels to safely image human subjects. We conducted all experiments and imaging sessions in a controlled radiofrequency safety lab and the whole-body 10.5T scanner in the Center for Magnetic Resonance Research. RESULTS: Quantitative agreement between the simulated and experimental results was obtained including S-parameters, B1+ maps, and SAR. We calculated peak 10 g average SAR using 4 different realistic human body models for a quadrature excitation and demonstrated that the peak 10 g SAR variation between subjects was less than 30%. We calculated safe power limits based on this set and used those limits to acquire T2 - and T2∗ -weighted images of human subjects at 10.5T. CONCLUSIONS: In this study, we acquired the first in vivo human head images at 10.5T using an 8-channel transmit/receive coil. We implemented and expanded a previously proposed workflow to validate the electromagnetic simulation model of the 8-channel transmit/receive coil. Using the validated coil model, we calculated radiofrequency power levels to safely image human subjects.


Assuntos
Imageamento por Ressonância Magnética , Ondas de Rádio , Simulação por Computador , Humanos , Imagens de Fantasmas
16.
Magn Reson Med ; 84(1): 289-303, 2020 07.
Artigo em Inglês | MEDLINE | ID: mdl-31846121

RESUMO

PURPOSE: To investigate the feasibility of imaging the human torso and to evaluate the performance of several radiofrequency (RF) management strategies at 10.5T. METHODS: Healthy volunteers were imaged on a 10.5T whole-body scanner in multiple target anatomies, including the prostate, hip, kidney, liver, and heart. Phase-only shimming and spoke pulses were used to demonstrate their performance in managing the B1+ inhomogeneity present at 447 MHz. Imaging protocols included both qualitative and quantitative acquisitions to show the feasibility of imaging with different contrasts. RESULTS: High-quality images were acquired and demonstrated excellent overall contrast and signal-to-noise ratio. The experimental results matched well with predictions and suggested good translational capabilities of the RF management strategies previously developed at 7T. Phase-only shimming provided increased efficiency, but showed pronounced limitations in homogeneity, demonstrating the need for the increased degrees of freedom made possible through single- and multispoke RF pulse design. CONCLUSION: The first in-vivo human imaging was successfully performed at 10.5T using previously developed RF management strategies. Further improvement in RF coils, transmit chain, and full integration of parallel transmit functionality are needed to fully realize the benefits of 10.5T.


Assuntos
Interpretação de Imagem Assistida por Computador , Imageamento por Ressonância Magnética , Coração , Humanos , Masculino , Ondas de Rádio , Razão Sinal-Ruído
17.
Magn Reson Med ; 82(1): 495-509, 2019 07.
Artigo em Inglês | MEDLINE | ID: mdl-30803023

RESUMO

PURPOSE: Despite the clear synergy between high channel counts in a receive array and magnetic fields ≥ 7 Tesla, to date such systems have been restricted to a maximum of 32 channels. Here, we examine SNR gains at 7 Tesla in unaccelerated and accelerated images with a 64-receive channel (64Rx) RF coil. METHODS: A 64Rx coil was built using circular loops tiled in 2 separable sections of a close-fitting form; custom designed preamplifier boards were integrated into each coil element. A 16-channel transmitter arranged in 2 rows along the z-axis was employed. The performance of the 64Rx array was experimentally compared to that of an industry-standard 32-channel receive (32Rx) array for SNR in unaccelerated images and for noise amplification under parallel imaging. RESULTS: SNR gains were observed in the periphery but not in the center of the brain in unaccelerated imaging compared to the 32Rx coil. With either 1D or 2D undersampling of k-space, or with slice acceleration together with 1D undersampling of k-space, significant reductions in g-factor noise were observed throughout the brain, yielding effective gains in SNR in the entire brain compared to the 32Rx coil. Task-based FMRI data with 12-fold 2D (slice and phase-encode) acceleration yielded excellent quality functional maps with the 64Rx coil but was significantly beyond the capabilities of the 32Rx coil. CONCLUSION: The results confirm the expectations from modeling studies and demonstrate that whole-brain studies with up to 16-fold, 2D acceleration would be feasible with the 64Rx coil.


Assuntos
Encéfalo/diagnóstico por imagem , Processamento de Imagem Assistida por Computador/instrumentação , Processamento de Imagem Assistida por Computador/métodos , Imageamento por Ressonância Magnética/métodos , Neuroimagem/métodos , Adulto , Desenho de Equipamento , Feminino , Humanos , Masculino , Imagens de Fantasmas , Razão Sinal-Ruído , Adulto Jovem
18.
Magn Reson Med ; 79(1): 511-514, 2018 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-28342176

RESUMO

PURPOSE: In this work, we investigated the relative effects of static magnetic field exposure (10.5 Tesla [T]) on two physiological parameters; blood pressure (BP) and heart rate (HR). METHODS: In vivo, we recorded both BP and HR in 4 swine (3 female, 1 male) while they were positioned within a 10.5T magnet. All measurements were performed invasively within these anesthetized animals by the placement of pressure catheters into their carotid arteries. RESULTS: We measured average increases of 2.0 mm Hg (standard deviation [SD], 6.9) in systolic BP and an increase of 4.5 mm Hg (SD, 13.7) in the diastolic BPs: We also noted an average increase of 1.2 beats per minute (SD, 2.5) in the HRs during such. CONCLUSION: Data regarding changes in BP and HR in anesthetized swine attributed to whole-body 10.5T exposure are reported. Magn Reson Med 79:511-514, 2018. © 2017 International Society for Magnetic Resonance in Medicine.


Assuntos
Anestesia , Pressão Sanguínea , Frequência Cardíaca , Campos Magnéticos , Animais , Determinação da Pressão Arterial , Artérias Carótidas/diagnóstico por imagem , Diástole , Feminino , Imageamento por Ressonância Magnética , Masculino , Suínos , Sístole
19.
Magn Reson Med ; 79(1): 479-488, 2018 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-28370375

RESUMO

PURPOSE: To validate electromagnetic and thermal simulations with in vivo temperature measurements, and to demonstrate a framework that can be used to predict temperature increase caused by radiofrequency (RF) excitation with dipole transmitter arrays. METHODS: Dipole arrays were used to deliver RF energy in the back/neck region of the swine using different RF excitation patterns (n = 2-4 per swine) for heating. The temperature in anesthetized swine (n = 3) was measured using fluoroscopic probes (n = 12) and compared against thermal modeling from animal-specific electromagnetic simulations. RESULTS: Simulated temperature curves were in agreement with the measured data. The root mean square error between simulated and measured temperature rise at all locations (at the end of each RF excitation) is calculated as 0.37°C. The mean experimental temperature rise at the maximum temperature rise locations (averaged over all experiments) is calculated as 2.89°C. The root mean square error between simulated and measured temperature at the maximum temperature rise location is calculated as 0.57°C. (Error values are averaged over all experiments.) CONCLUSIONS: Electromagnetic and thermal simulations were validated with experiments. Thermal effects of RF excitation at 10.5 Tesla with dipoles were investigated. Magn Reson Med 79:479-488, 2018. © 2017 International Society for Magnetic Resonance in Medicine.


Assuntos
Desenho de Equipamento , Temperatura Alta , Hipertermia Induzida/instrumentação , Ondas de Rádio , Animais , Calibragem , Simulação por Computador , Campos Eletromagnéticos , Radiação Eletromagnética , Imageamento por Ressonância Magnética , Espectroscopia de Ressonância Magnética , Modelos Anatômicos , Imagens de Fantasmas , Suínos , Tomografia Computadorizada por Raios X
20.
Magn Reson Med ; 77(1): 434-443, 2017 01.
Artigo em Inglês | MEDLINE | ID: mdl-27770469

RESUMO

PURPOSE: To explore the potential of performing body imaging at 10.5 Tesla (T) compared with 7.0T through evaluating the transmit/receive performance of similarly configured dipole antenna arrays. METHODS: Fractionated dipole antenna elements for 10.5T body imaging were designed and evaluated using numerical simulations. Transmit performance of antenna arrays inside the prostate, kidneys and heart were investigated and compared with those at 7.0T using both phase-only radiofrequency (RF) shimming and multi-spoke pulses. Signal-to-noise ratio (SNR) comparisons were also performed. A 10-channel antenna array was constructed to image the abdomen of a swine at 10.5T. Numerical methods were validated with phantom studies at both field strengths. RESULTS: Similar power efficiencies were observed inside target organs with phase-only shimming, but RF nonuniformity was significantly higher at 10.5T. Spokes RF pulses allowed similar transmit performance with accompanying local specific absorption rate increases of 25-90% compared with 7.0T. Relative SNR gains inside the target anatomies were calculated to be >two-fold higher at 10.5T, and 2.2-fold SNR gain was measured in a phantom. Gradient echo and fast spin echo imaging demonstrated the feasibility of body imaging at 10.5T with the designed array. CONCLUSION: While comparable power efficiencies can be achieved using dipole antenna arrays with static shimming at 10.5T; increasing RF nonuniformities underscore the need for efficient, robust, and safe parallel transmission methods. Magn Reson Med 77:434-443, 2017. © 2016 International Society for Magnetic Resonance in Medicine.


Assuntos
Simulação por Computador , Processamento de Imagem Assistida por Computador/métodos , Imageamento por Ressonância Magnética/instrumentação , Imageamento por Ressonância Magnética/métodos , Imagem Corporal Total/métodos , Adulto , Animais , Desenho de Equipamento , Feminino , Humanos , Masculino , Modelos Anatômicos , Imagens de Fantasmas , Suínos