Biomed Tech (Berl) ; 34(3): 53-6, 1989 Mar.
Artigo em Alemão | MEDLINE | ID: mdl-2720047

RESUMO

To date, analysis of temperature in the drill area of cortical bone have been limited to measurements with thermocouple systems at a certain distance from the drill hole. Many authors equate this temperature measurement with the drill--cortical bone interface temperature. In order show that there is a temperature difference, a drill hole was simulated with the aid of the "Finite Element Method". The interface temperature was calculated by the energy distribution. It was shown, that for "dry" and "watercooled" drilling, the drill hole temperature was 13 degrees C higher than the temperature measured with the thermocouple systems at a distance 0.5 mm of from the drill hole. In particular when using "watercooled" drills for bone and dental surgery, the temperature may be higher than the bone damage limit of 44 degrees C for lengthy and 50 degrees C for brief drilling.


Assuntos
Regulação da Temperatura Corporal , Osso e Ossos/cirurgia , Instrumentos Cirúrgicos , Gráficos por Computador , Humanos , Software