Med Image Comput Comput Assist Interv ; 14(Pt 1): 379-86, 2011.
Artigo em Inglês | MEDLINE | ID: mdl-22003640

RESUMO

PURPOSE: This paper is the first report on the monitoring of tissue ablation using ultrasound RF echo time series. METHODS: We calcuate frequency and time domain features of time series of RF echoes from stationary tissue and transducer, and correlate them with ablated and non-ablated tissue properties. RESULTS: We combine these features in a nonlinear classification framework and demonstrate up to 99% classification accuracy in distinguishing ablated and non-ablated regions of tissue, in areas as small as 12mm2 in size. We also demonstrate significant improvement of ablated tissue classification using RF time series compared to the conventional approach of using single RF scan lines. CONCLUSIONS: The results of this study suggest RF echo time series as a promising approach for monitoring ablation, and capturing the changes in the tissue microstructure as a result of heat-induced necrosis.


Assuntos
Ablação por Cateter/instrumentação , Ondas de Choque de Alta Energia , Interpretação de Imagem Assistida por Computador/métodos , Algoritmos , Calibragem , Ablação por Cateter/métodos , Desenho de Equipamento , Temperatura Alta , Humanos , Necrose , Ondas de Rádio , Reprodutibilidade dos Testes , Fatores de Tempo , Ultrassom