Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Assunto principal
Ano de publicação
Tipo de documento
Assunto da revista
País de afiliação
Intervalo de ano de publicação
1.
Spectrochim Acta A Mol Biomol Spectrosc ; 310: 123895, 2024 Apr 05.
Artigo em Inglês | MEDLINE | ID: mdl-38262294

RESUMO

Using optical density at 600 nm (OD600) to measure the microbial concentration is a popular approach due to its advantages like quick response and non-destructive. However, the OD600 measurement might be affected by the metabolic pigment, and it would become invalid when the solution dilution is insufficient. To overcome these issues, we proposed to adopt a more robust wavelength at 890 nm to quantify the attenuation of transmission light. After selecting this light source, we designed the light path and the circuit of the online monitoring device. Meanwhile, the random forest algorithm was introduced for temperature compensation and improving the stability of the device. This device was verified by monitoring the microbial concentration of four strains (Yeast, Bacillus, Arthrobacter, and Escherichia coli). The experimental result suggested that the mean absolute percentage error reached 4.11 %, 4.28 %, 4.49 %, and 4.53 % respectively, which is helpful to improve the accuracy of microbial concentration measurement.


Assuntos
Bacillus , Escherichia coli/metabolismo , Temperatura
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA