1.
Analyst ; 136(1): 120-7, 2011 Jan 07.
Artigo em Inglês | MEDLINE | ID: mdl-20976316

RESUMO

Gaseous ammonia, a common toxic industrial compound, is not detected readily in ion trap mass spectrometers because its molecular ion falls below the low-mass cutoff (~m/z 40) normally used when examining organic compounds. Instead, reactions of ammonia with halobenzene radical cations were used with internal electron ionization in two cylindrical ion trap miniature mass spectrometers to create a characteristic product ion by which to identify and quantify ammonia. Ammonia showed a linear response over the concentration range studied (parts per million [ppm] to parts per billion [ppb]) with limits of detection of 17 ppm and 220 ppb for experiments involving direct introduction and thermal desorption after pre-concentration, respectively. These values are comparable to ammonia's permissible exposure limit (50 ppm) and odor threshold (5 ppm). Receiver operating characteristic (ROC) curves were used to describe the method sensitivity, the probability of true positives, and the false positive rate for ammonia. A customized reaction scan function was created to select the species available for the ion/molecule reaction and set the amount of time the product ion could be accumulated in the trap. Product ion identity was verified using tandem mass spectrometry. Similar reactions with methylamine, ethylamine and the two nitriles, acetonitrile and benzonitrile, were explored.


Assuntos
Amônia/análise , Íons/química , Espectrometria de Massas em Tandem/métodos , Aminas/química , Bromobenzenos/química , Clorobenzenos/química , Miniaturização , Nitrilas/química
2.
J Am Soc Mass Spectrom ; 32(2): 509-518, 2021 Feb 03.
Artigo em Inglês | MEDLINE | ID: mdl-33382610

RESUMO

Cycloidal sector mass analyzers have, in principle, perfect focusing due to perpendicularly oriented uniform electric and magnetic fields, making them ideal candidates for incorporation of spatially coded apertures. We have previously demonstrated a proof-of-concept cycloidal-coded aperture miniature mass spectrometer (C-CAMMS) instrument and achieved a greater than 10-fold increase in throughput without sacrificing resolution, compared with a single slit instrument. However, artifacts were observed in the reconstructed mass spectrum due to nonuniformity in the electric field and misalignment of the detector and the ion source with the mass analyzer focal plane. In this work, we modified the mass analyzer design of the previous C-CAMMS instrument to improve electric field uniformity, improve the alignment of the ion source and the mass analyzer with the detector, and increase the depth-of-focus to further facilitate alignment. A comparison of reconstructed spectra of a mixture of dry air and toluene at different electric fields was performed using the improved C-CAMMS prototype. A reduction in reconstruction artifacts compared to our proof-of-concept C-CAMMS instrument highlights the improved performance enabled by the design changes.

3.
J Am Soc Mass Spectrom ; 19(10): 1419-24, 2008 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-18703351

RESUMO

A recently developed prototype mobile laboratory mass spectrometer, incorporating an atmospheric pressure ionization (API) interface, is described. This system takes advantage of the small size, lower voltage requirements, and tandem MS abilities of the cylindrical ion trap mass analyzer. The prototype API MS uses small, low-power pumps to fit into a 0.1-m(3) self-contained package weighing <45 kg. This instrument has been adapted to allow rapid interfacing to electrospray ionization, desorption electrospray ionization, and direct analysis in real-time sources. Initial data indicate that these techniques provide rapid detection and identification of compounds for quality control, homeland security, and forensic applications. In addition, this instrument is self-contained and compact, making it ideally extensible to mobile laboratory and field analyses. Initial MS and MS/MS data for analyses of drugs, food, and explosives are presented herein.

4.
J Am Soc Mass Spectrom ; 29(2): 360-372, 2018 02.
Artigo em Inglês | MEDLINE | ID: mdl-29052038

RESUMO

Despite many potential applications, miniature mass spectrometers have had limited adoption in the field due to the tradeoff between throughput and resolution that limits their performance relative to laboratory instruments. Recently, a solution to this tradeoff has been demonstrated by using spatially coded apertures in magnetic sector mass spectrometers, enabling throughput and signal-to-background improvements of greater than an order of magnitude with no loss of resolution. This paper describes a proof of concept demonstration of a cycloidal coded aperture miniature mass spectrometer (C-CAMMS) demonstrating use of spatially coded apertures in a cycloidal sector mass analyzer for the first time. C-CAMMS also incorporates a miniature carbon nanotube (CNT) field emission electron ionization source and a capacitive transimpedance amplifier (CTIA) ion array detector. Results confirm the cycloidal mass analyzer's compatibility with aperture coding. A >10× increase in throughput was achieved without loss of resolution compared with a single slit instrument. Several areas where additional improvement can be realized are identified. Graphical Abstract ᅟ.

5.
J Am Soc Mass Spectrom ; 26(11): 1903-10, 2015 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-26133527

RESUMO

A novel chemical ionization (CI) source has been developed based on a carbon nanotube (CNT) field emission electron source. The CNT-based electron source was evaluated and compared with a standard filament thermionic electron source in a commercial explosives trace detection desktop mass spectrometer. This work demonstrates the first reported use of a CNT-based ion source capable of collecting CI mass spectra. Both positive and negative modes were investigated. Spectra were collected for a standard mass spectrometer calibration compound, perfluorotributylamine (PFTBA), as well as trace explosives including trinitrotoluene (TNT), Research Department explosive (RDX), and pentaerythritol tetranitrate (PETN). The electrical characteristics, lifetime at operating pressure, and power requirements of the CNT-based electron source are reported. The CNT field emission electron sources demonstrated an average lifetime of 320 h when operated in constant emission mode under elevated CI pressures. The ability of the CNT field emission source to cycle on and off can provide enhanced lifetime and reduced power consumption without sacrificing performance and detection capabilities. Graphical Abstract ᅟ.