RESUMO
OBJECTIVES: Quantitative human chorionic gonadotropin (hCG) measurements are used to manage women classified with a pregnancy of unknown location (PUL). Two point of care testing (POCT) devices that quantify hCG are commercially available. We verified the i-STAT 1 (Abbott) and the AQT 90 FLEX (Radiometer) prior to use in PUL triage. METHODS: Tests for precision, external quality assurance (EQA), correlation, hook effect and recovery were undertaken alongside a POCT usability assessment during this prospective multi-center verification. RESULTS: Coefficients of variation ranged between 4.0 and 5.1â¯% for the three i-STAT 1 internal quality control (IQC) solutions and between 6.8 and 7.3â¯% for the two AQT IQC solutions. Symmetric differences in POCT EQA results when compared with laboratory and EQA stock values ranged between 3.2 and 24.5â¯% for the i-STAT 1 and between 3.3 and 36.9â¯% for the AQT. Correlation coefficients (i-STAT 1: 0.96, AQT: 0.99) and goodness of fit curves (i-STAT 1: 0.92, AQT: 0.99) were excellent when using suitable whole blood samples. An hCG hook effect was noted with the i-STAT 1 between 572,194 and 799,089â¯IU/L, lower than the hook effect noted with the AQT, which was between 799,089 and 1,619,309â¯IU/L. When hematocrit concentration was considered in sample types validated for use with each device, hCG recovery was 108â¯% with the i-STAT 1 and 98â¯% with the AQT. The i-STAT 1 scored lower on usability overall (90/130) than the AQT (121/130, p<0.001, Mann-Whitney). CONCLUSIONS: Both hCG POCT devices were verified for use in clinical practice. Practical factors must also be considered when choosing which device to use in each unit.