1.
Opt Lett ; 33(12): 1333-5, 2008 Jun 15.
Artigo em Inglês | MEDLINE | ID: mdl-18552949

RESUMO

A fast handheld two-photon fiber-optic fluorescence endoscope for three-dimensional (3D) in vivo cellular imaging is developed. The compact handheld probe of the two-photon endoscope can simply be placed into contact with the target tissue to reveal clear 3D surface and subsurface histological images without biopsy. The new system has, to the best of our knowledge, the largest field of view (FOV) of 475 microm x 475 microm and a 3D imaging volume larger than 475 microm x 475 microm x 250 microm. A real-time two-photon fluorescence image is displayed at 0.4 mm(2)/s. The lateral and axial resolutions of the two-photon fluorescence endoscope are better than 1 and 14.5 microm, respectively.


Assuntos
Rim/ultraestrutura , Animais , Endoscopia/métodos , Tecnologia de Fibra Óptica/instrumentação , Tecnologia de Fibra Óptica/métodos , Microscopia de Fluorescência por Excitação Multifotônica/instrumentação , Microscopia de Fluorescência por Excitação Multifotônica/métodos , Fibras Ópticas , Ratos