1.
Acta Orthop ; 93: 444-450, 2022 04 19.
Artigo em Inglês | MEDLINE | ID: mdl-35438184

RESUMO

BACKGROUND AND PURPOSE: We have previously reported that the whole blood (WB) chromium (Cr) and cobalt (Co) ion levels decrease in the short term after ReCap-M2a-Magnum large-diameter head (LDH) metal-on-metal (MoM) total hip arthroplasty (THA). This study reports long-term metal ion levels and clinical outcomes after ReCap-Magnum THA. PATIENTS AND METHODS: ReCap-M2a-Magnum LDH THA was used in 1,450 patients in our hospital district from 2005 to 2012. Median follow-up time was 10 years. 991 patients had 2 or more metal ion measurements. The median measurement interval was 4 years. Individual metal ion change was assessed using logarithmic metal ion values in a random coefficient model. Kaplan-Meier survival estimates were calculated for revision surgery for any reason for revision, and separately for metal-related adverse events (metal ions above safe upper limit [SUL], revision due to ARMD, or pseudotumor). RESULTS: Geometric mean of Cr decreased from 1.8 ppb (geometric standard deviation [GSD] 1.8) to 1.0 ppb (GSD 2.8, p < 0.001). The Co levels decreased from 1.7 ppb (GSD 2.4) to 1.4 ppb (GSD 2.8, p < 0.001). The hip-specific survival was 85% for revision due to any reason at 14 years and the hip-specific survival for any metal-related adverse event was 69% at 14 years. INTERPRETATION: WB Cr and Co levels continued to decrease in the long-term follow-up of ReCap-M2a-Magnum THA patients. The amount of metal-related adverse events was rather high, but revision surgery was seldom required. We suggest that after 10 years from the implantation a 5-year measurement interval may be sufficient for asymptomatic ReCap-M2a-Magnum patients.


Assuntos
Artroplastia de Quadril , Prótese de Quadril , Próteses Articulares Metal-Metal , Artroplastia de Quadril/efeitos adversos , Cromo , Cobalto , Prótese de Quadril/efeitos adversos , Humanos , Íons , Próteses Articulares Metal-Metal/efeitos adversos , Metais , Desenho de Prótese , Falha de Prótese , Reoperação
2.
Acta Orthop ; 93: 241-248, 2022 01 19.
Artigo em Inglês | MEDLINE | ID: mdl-35043969

RESUMO

Background and purpose - Data regarding long-term behavior of metal ion levels in metal-on-metal total hip arthroplasty (MoM THA) patients is scarce. Therefore, we assessed whether there is any change in whole blood (WB) chromium (Cr), and cobalt (Co) ion measurements in Durom and MMC MoM THA patients over time. The secondary aim was to report the clinical outcomes using these devices in a single district. Patients and methods - Durom and MMC cups were used in 249 MoM THAs from 2005 to 2011 in our district. Median follow-up time was 12 years for Durom THA (interquartile range [IQR] = 3) and 9 years for MMC THA (IQR = 1). A random coefficient model was used to compare individual differences in repeated WB Cr and Co ion measurements. The Kaplan-Meier estimator was used to analyze implant survival with any reason for revision as the endpoint. Results - Geometric means of Cr in Durom THA and MMC THA patients decreased from 2.2 ppb (geometric standard deviation [SD] = 1.9) to 1.5 ppb (geometric SD = 2.5, p< 0.001) and from 1.8 ppb (geometric SD = 1.8) to 1.1 ppb (geometric SD = 2.8, p = 0.01) respectively. The geometric means of Co values remained unchanged. The 10-year survival of Durom THA was 82%, and that of MMC THA 89% for any revision reason as endpoint. Interpretation - WB Cr levels decreased over time, and Co levels remained unchanged at long-term follow-up. Despite this we recommend continuing the follow-up of these devices due to relatively low implant survival.


Assuntos
Artroplastia de Quadril , Prótese de Quadril , Próteses Articulares Metal-Metal , Artroplastia de Quadril/métodos , Cromo , Cobalto , Prótese de Quadril/efeitos adversos , Humanos , Próteses Articulares Metal-Metal/efeitos adversos , Metais , Desenho de Prótese
3.
Am J Emerg Med ; 50: 602-605, 2021 12.
Artigo em Inglês | MEDLINE | ID: mdl-34592568

RESUMO

This brief report is a retrospective review of three cases of iatrogenic digital ischemia and clinical outcome at six months. Hand injuries are one of the most common injuries that occur in the working population. Iatrogenic digital ischemia is a rare condition that can be avoided by proper wound management. After the correct initial treatment is provided, it is important to apply the wound dressing correctly to avoid iatrogenic trauma or ischemia. Currently, there is no consensus regarding the best treatment for these injuries. Our aim is to remind clinicians of this rare condition, and to highlight prevention and treatment strategies.


Assuntos
Serviço Hospitalar de Emergência , Traumatismos dos Dedos/etiologia , Traumatismos dos Dedos/patologia , Dedos/irrigação sanguínea , Isquemia/etiologia , Isquemia/patologia , Adulto , Feminino , Traumatismos dos Dedos/terapia , Humanos , Isquemia/terapia , Masculino
4.
Acta Orthop ; 91(4): 378-382, 2020 08.
Artigo em Inglês | MEDLINE | ID: mdl-32285731

RESUMO

Background and purpose - Whole-blood (WB) chromium (Cr) and cobalt (Co) measurements are vital in the follow-up of metal-on-metal total hip replacement (MoM THR) patients. We examined whether there is a substantial change in repeated WB, Co, and Cr levels in patients with bilateral ReCap-M2A-Magnum THR. We also specified the number of patients exceeding the safe upper limit (SUL) of WB Co and Cr in the repeated measurement.Patients and methods - We identified 141 patients with bilateral ReCap-M2A-Magnum THR operated in our institution. 61 patients had repeated WB metal ion measurements with bilateral MoM implants still in situ in the second measurement. The mean time elapsing from the first measurement (initial measurement) to the second (control measurement) was 1.9 years (SD = 0.6, range 0.2-3.5). We used earlier established SUL levels for bilateral implants by Van Der Straeten et al. (2013).Results - The median (range) Co and Cr values decreased in the repeated measurement from 2.7 (0.6-25) to 2.1 (0.5-21) and 2.6 (0.8-14) to 2.1 (0.5-18) respectively. In 13% of the patients Co levels exceeded the SUL in the initial measurement and the proportion remained constant, at 13%, in the repeated measurement. In 5% of the patients, Cr levels were above SUL in the initial measurement and an equal 5% in the control measurement.Interpretation - Repeated WB metal ion levels did not increase in patients with bilateral ReCap-M2A-Magnum THR with a mean 1.9-year measurement interval. Long-term development of WB metal ion levels is still unclear in these patients.


Assuntos
Artroplastia de Quadril , Cromo/sangue , Cobalto/sangue , Prótese de Quadril , Próteses Articulares Metal-Metal , Artroplastia de Quadril/efeitos adversos , Artroplastia de Quadril/estatística & dados numéricos , Feminino , Prótese de Quadril/efeitos adversos , Prótese de Quadril/estatística & dados numéricos , Humanos , Masculino , Próteses Articulares Metal-Metal/efeitos adversos , Próteses Articulares Metal-Metal/estatística & dados numéricos , Pessoa de Meia-Idade , Fatores de Tempo
5.
Scand J Surg ; 111(1): 14574969211066197, 2022.
Artigo em Inglês | MEDLINE | ID: mdl-35168442

RESUMO

BACKGROUND AND OBJECTIVE: Our aim was to assess long-term metal ion level changes and clinical outcome in patients with a Birmingham hip arthroplasty. METHODS: For the purpose of this study, we identified all BHR hip resurfacing arthroplasty (HRA) and total hip arthroplasty (THA) operations performed in Turku University Hospital. A random coefficient model was used to compare the change between the first and last metal ion measurement. A Kaplan-Meier estimator was used to assess the survivorship of the BHR HRA and BHR THA with metal related adverse events (pseudotumor, elevated metal ions above the safe upper limit, revision due to metallosis), or revision due to any reason as endpoints with 95% confidence intervals (CIs). RESULTS: BHR HRA was used in 274 hips (233 patients). In addition, we identified 38 BHR-Synergy THAs (38 patients). Operations were performed between 2003 and 2010. Median follow-up time was 14 years for BHR HRA (range: 0.6-17) and 11 years for BHR THA (range: 4.7-13). In the BHR HRA group, geometric means of Cr and Co levels decreased from 2.1 to 1.6 ppb and 2.4 to 1.5 ppb, respectively, during a 3.0-year measurement interval. Metal ion levels in the BHR THA group did not show notable increase. The survivorship of BHR HRA was 66% in 16 years and 34% for BHR THA at 12 years for any metal-related adverse event. CONCLUSIONS: Patients with a Birmingham hip device do not seem to benefit from frequent repeated metal ion measurements. The amount of patients with metal-related adverse events was relatively high, but many of them did not require surgery.