Clin Oral Implants Res ; 32(11): 1288-1298, 2021 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-34352139

RESUMO

OBJECTIVE: To evaluate the influence of artificial aging on the transformation propagation and fracture resistance of zirconia implants. METHODS: One-piece (with integrated implant abutment, 1P; regular diameter [4.1mm]; n = 16) and two-piece (with separate implant abutment, 2P; wide diameter [5 mm]; n = 16) zirconia implants were embedded according to ISO 14801. A two-piece titanium-zirconium implant (Ti-Zr; 4.1 mm diameter) served as a control (n = 16). One subgroup (n = 8) of each system was simultaneously dynamically loaded (107 cycles; 98N) and hydrothermally aged (85°C, 58 days), while the other subgroup (n = 8) remained untreated. Finally, specimens were statically loaded to fracture. Potential crystal phase transformation was examined at cross sections using scanning electron microscopy (SEM). A multivariate linear regression model was applied for statistical analyses. RESULTS: The fracture resistance of 1P (1,117 [SD = 38] N; loaded/aged: 1,009 [60] N), 2P (850 [36] N; loaded/aged: 799 [84] N), and Ti-Zr implants (1,338 [205] N; loaded/aged: 1,319 [247] N) was not affected significantly by loading/aging (p = .171). However, when comparing the systems, they revealed significant differences independent of loading/aging (p ≤ .001). Regarding the crystal structure, a transformation zone was observed in SEM images of 1P only after aging, while 2P showed a transformation zone even before aging. After hydrothermal treatment, an increase of this monoclinic layer was observed in both systems. CONCLUSIONS: The Ti-Zr control implant showed higher fracture resistance compared to both zirconia implants. Loading/aging had no significant impact on the fracture resistance of both zirconia implants. The wide-body 2P zirconia implant was weaker than the regular body 1P implant.


Assuntos
Implantes Dentários , Zircônio , Dente Suporte , Projeto do Implante Dentário-Pivô , Falha de Restauração Dentária , Análise do Estresse Dentário , Teste de Materiais , Titânio
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA