Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 3 de 3
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
Artif Organs ; 40(9): 894-903, 2016 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-26748566

RESUMO

Preventing ventricular suction and venous congestion through balancing flow rates and circulatory volumes with dual rotary ventricular assist devices (VADs) configured for biventricular support is clinically challenging due to their low preload and high afterload sensitivities relative to the natural heart. This study presents the in vivo evaluation of several physiological control systems, which aim to prevent ventricular suction and venous congestion. The control systems included a sensor-based, master/slave (MS) controller that altered left and right VAD speed based on pressure and flow; a sensor-less compliant inflow cannula (IC), which altered inlet resistance and, therefore, pump flow based on preload; a sensor-less compliant outflow cannula (OC) on the right VAD, which altered outlet resistance and thus pump flow based on afterload; and a combined controller, which incorporated the MS controller, compliant IC, and compliant OC. Each control system was evaluated in vivo under step increases in systemic (SVR ∼1400-2400 dyne/s/cm(5) ) and pulmonary (PVR ∼200-1000 dyne/s/cm(5) ) vascular resistances in four sheep supported by dual rotary VADs in a biventricular assist configuration. Constant speed support was also evaluated for comparison and resulted in suction events during all resistance increases and pulmonary congestion during SVR increases. The MS controller reduced suction events and prevented congestion through an initial sharp reduction in pump flow followed by a gradual return to baseline (5.0 L/min). The compliant IC prevented suction events; however, reduced pump flows and pulmonary congestion were noted during the SVR increase. The compliant OC maintained pump flow close to baseline (5.0 L/min) and prevented suction and congestion during PVR increases. The combined controller responded similarly to the MS controller to prevent suction and congestion events in all cases while providing a backup system in the event of single controller failure.


Assuntos
Insuficiência Cardíaca/terapia , Ventrículos do Coração/cirurgia , Coração Auxiliar , Animais , Desenho de Equipamento , Feminino , Insuficiência Cardíaca/fisiopatologia , Insuficiência Cardíaca/cirurgia , Ventrículos do Coração/fisiopatologia , Hemodinâmica , Modelos Cardiovasculares , Circulação Pulmonar , Ovinos , Resistência Vascular , Função Ventricular Esquerda , Função Ventricular Direita
2.
Artif Organs ; 39(2): 102-9, 2015 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-25041754

RESUMO

Biventricular support with dual rotary ventricular assist devices (VADs) has been implemented clinically with restriction of the right VAD (RVAD) outflow cannula to artificially increase afterload and, therefore, operate within recommended design speed ranges. However, the low preload and high afterload sensitivity of these devices increase the susceptibility of suction events. Active control systems are prone to sensor drift or inaccurate inferred (sensor-less) data, therefore an alternative solution may be of benefit. This study presents the in vitro evaluation of a compliant outflow cannula designed to passively decrease the afterload sensitivity of rotary RVADs and minimize left-sided suction events. A one-way fluid-structure interaction model was initially used to produce a design with suitable flow dynamics and radial deformation. The resultant geometry was cast with different initial cross-sectional restrictions and concentrations of a softening diluent before evaluation in a mock circulation loop. Pulmonary vascular resistance (PVR) was increased from 50 dyne s/cm(5) until left-sided suction events occurred with each compliant cannula and a rigid, 4.5 mm diameter outflow cannula for comparison. Early suction events (PVR ∼ 300 dyne s/cm(5) ) were observed with the rigid outflow cannula. Addition of the compliant section with an initial 3 mm diameter restriction and 10% diluent expanded the outflow restriction as PVR increased, thus increasing RVAD flow rate and preventing left-sided suction events at PVR levels beyond 1000 dyne s/cm(5) . Therefore, the compliant, restricted outflow cannula provided a passive control system to assist in the prevention of suction events with rotary biventricular support while maintaining pump speeds within normal ranges of operation.


Assuntos
Catéteres , Coração Auxiliar , Desenho de Equipamento , Hemodinâmica , Humanos , Modelos Cardiovasculares , Sucção
3.
Ann Biomed Eng ; 44(5): 1370-80, 2016 May.
Artigo em Inglês | MEDLINE | ID: mdl-26283049

RESUMO

The low preload and high afterload sensitivities of rotary ventricular assist devices (VADs) may cause ventricular suction events or venous congestion. This is particularly problematic with rotary biventricular support (BiVAD), where the Starling response is diminished in both ventricles. Therefore, VADs may benefit from physiological control systems to prevent adverse events. This study compares active, passive and combined physiological controllers for rotary BiVAD support with constant speed mode. Systemic (SVR) and pulmonary (PVR) vascular resistance changes and exercise were simulated in a mock circulation loop to evaluate the capacity of each controller to prevent suction and congestion and increase exercise capacity. All controllers prevented suction and congestion at high levels of PVR (900 dynes s cm(-5)) and SVR (3000 dynes s cm(-5)), however these events occurred in constant speed mode. The controllers increased preload sensitivity (0.198-0.34 L min(-1) mmHg(-1)) and reduced afterload sensitivity (0.0001-0.008 L min(-1) mmHg(-1)) of the VADs when compared to constant speed mode (0.091 and 0.072 L min(-1) mmHg(-1) respectively). The active controller increased pump speeds (400-800 rpm) and pump flow by 2.8 L min(-1) during exercise, thus increasing exercise capacity. By reducing suction and congestion and by increasing exercise capacity, the control systems presented in this study may help increase quality of life of VAD patients.


Assuntos
Coração Auxiliar , Modelos Cardiovasculares , Resistência Vascular , Humanos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA