RESUMO
Recently, the Heart Rhythm Society published recommendations on management of patients with cardiac implantable electronic device (CIED) who require radiotherapy (RT). We aimed to report the experience of a teaching hospital, and discuss our practice in the context of recently published guidelines. We identified all consecutive CIED recipients (12,736 patients) who underwent RT between March 2006 and June 2017. Among them, 90 (1%) patients (78.2 ± 10 years, 73% male) had a CIED: 82 pacemakers and 8 implantable cardioverter-defibrillators. Two patients required CIED extraction prior to RT for ipsilateral breast cancer (no device replacement in 1 patient). Four patients (5%) were considered at high-risk, 35 (39%) at intermediate-risk, and the remaining 50 (56%) at low-risk for CIED dysfunction. Overall, only a minority of patients followed recommended local protocol during RT delivery (31%) and during follow-up (56%). CIED malfunction was detected in 5 patients (6%), mainly back-up mode resetting (80%), with 4 (including 3 pelvic cancer location) patients initially classified as being at intermediate-risk and 1 at low-risk. Four out of the 5 patients with CEID malfunction had received neutron producing beams. In conclusion, our findings underline the lack of rigorous monitoring of patients undergoing RT (though CIED malfunction appears to be rare and relatively benign in nature), and emphasize the interest of considering neutron producing beam for risk stratification as recommended in recent guidelines. Optimization of patient's management requires a close collaboration between both CIED clinicians and radiation oncologists, and more systematic remote CIED monitoring may be helpful.