Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
País/Região como assunto
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Diabetologia ; 67(7): 1235-1244, 2024 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-38634887

RESUMO

AIMS/HYPOTHESIS: This study compares the efficacy and safety of a tubeless, on-body automated insulin delivery (AID) system with that of a tubeless, on-body sensor-augmented pump (SAP). METHODS: This multicentre, parallel-group, RCT was conducted at 13 tertiary medical centres in South Korea. Adults aged 19-69 years with type 1 diabetes who had HbA1c levels of <85.8 mmol/mol (<10.0%) were eligible. The participants were assigned at a 1:1 ratio to receive a tubeless, on-body AID system (intervention group) or a tubeless, on-body SAP (control group) for 12 weeks. Stratified block randomisation was conducted by an independent statistician. Blinding was not possible due to the nature of the intervention. The primary outcome was the percentage of time in range (TIR), blood glucose between 3.9 and 10.0 mmol/l, as measured by continuous glucose monitoring. ANCOVAs were conducted with baseline values and study centres as covariates. RESULTS: A total of 104 participants underwent randomisation, with 53 in the intervention group and 51 in the control group. The mean (±SD) age of the participants was 40±11 years. The mean (±SD) TIR increased from 62.1±17.1% at baseline to 71.5±10.7% over the 12 week trial period in the intervention group and from 64.7±17.0% to 66.9±15.0% in the control group (difference between the adjusted means: 6.5% [95% CI 3.6%, 9.4%], p<0.001). Time below range, time above range, CV and mean glucose levels were also significantly better in the intervention group compared with the control group. HbA1c decreased from 50.9±9.9 mmol/mol (6.8±0.9%) at baseline to 45.9±7.4 mmol/mol (6.4±0.7%) after 12 weeks in the intervention group and from 48.7±9.1 mmol/mol (6.6±0.8%) to 45.7±7.5 mmol/mol (6.3±0.7%) in the control group (difference between the adjusted means: -0.7 mmol/mol [95% CI -2.0, 0.8 mmol/mol] (-0.1% [95% CI -0.2%, 0.1%]), p=0.366). No diabetic ketoacidosis or severe hypoglycaemia events occurred in either group. CONCLUSIONS/INTERPRETATION: The use of a tubeless, on-body AID system was safe and associated with superior glycaemic profiles, including TIR, time below range, time above range and CV, than the use of a tubeless, on-body SAP. TRIAL REGISTRATION: Clinical Research Information Service (CRIS) KCT0008398 FUNDING: The study was funded by a grant from the Korea Medical Device Development Fund supported by the Ministry of Science and ICT; the Ministry of Trade, Industry and Energy; the Ministry of Health and Welfare; and the Ministry of Food and Drug Safety (grant number: RS-2020-KD000056).


Assuntos
Glicemia , Diabetes Mellitus Tipo 1 , Hipoglicemiantes , Sistemas de Infusão de Insulina , Insulina , Humanos , Diabetes Mellitus Tipo 1/tratamento farmacológico , Diabetes Mellitus Tipo 1/sangue , Masculino , Pessoa de Meia-Idade , Adulto , Feminino , Insulina/administração & dosagem , Insulina/uso terapêutico , Glicemia/efeitos dos fármacos , Glicemia/metabolismo , Glicemia/análise , Hipoglicemiantes/administração & dosagem , Hipoglicemiantes/uso terapêutico , Idoso , Hemoglobinas Glicadas/metabolismo , Hemoglobinas Glicadas/análise , República da Coreia , Automonitorização da Glicemia/métodos , Adulto Jovem
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA