Clin Chem Lab Med ; 59(12): 1944-1949, 2021 11 25.
Artigo em Inglês | MEDLINE | ID: mdl-34473921

RESUMO

OBJECTIVES: In the clinical setting, the analysis and quantification of vitamin C (ascorbic acid) poses several challenges including analyte instability and poor retention by reverse phase HPLC systems. In this article we describe a rapid hydrophilic interaction chromatography ultraviolet method for the measurement of total vitamin C in plasma which overcomes these issues. METHODS: Ascorbic acid and the internal standard were separated under isocratic conditions using a Waters BEH-Amide column and a mobile phase containing 0.005 M potassium phosphate in 80% acetonitrile. RESULTS: The proposed method was validated and showed good precision (coefficient of variation <5%), accuracy (>99%), and analyte stability after extraction (>24 h). CONCLUSIONS: The simple sample preparation allows full automation and rapid analytical run times of the assay and is therefore suitable for a high-throughput clinical chromatography laboratory.


Assuntos
Ácido Ascórbico , Laboratórios , Cromatografia Líquida de Alta Pressão/métodos , Humanos , Reprodutibilidade dos Testes , Vitaminas