A comparison of point-of-care instruments designed for monitoring oral anticoagulation with standard laboratory methods.
Gosselin, R; Owings, J T; White, R H; Hutchinson, R; Branch, J; Mahackian, K; Johnston, M; Larkin, E C.
Afiliação
  • Gosselin R; University of California, Davis Medical Center, Sacramento, USA.
Thromb Haemost ; 83(5): 698-703, 2000 May.
Article em En | MEDLINE | ID: mdl-10823265
ABSTRACT
Our study compared point-of-care (POC) device monitoring with traditional clinical laboratory methods device of patients on oral anticoagulant therapy. The POC devices used in the study were Coumatrak, CoaguChek, CoaguChek Plus, Thrombolytic Assessment System (TAS) PT-One, TAS PTNC, TAS PT, Hemachron Jr. Signature, ProTime Microcoagulation System, and Medtronics ACT II. The clinical laboratory method used thromboplastins with different ISI values Innovin and Thromboplastin C Plus (TPC). All POC INRs showed strong correlation with both laboratory methods, with correlation coefficients of >0.900. All POC methods demonstrated a significant (p <0.05) difference in INR values, except the TAS PTNC and ACT II INRs (p 0.12 and 0.71 respectively) when compared with Innovin INRs. All POC INRs were significantly different from TPC generated INRs (p <0.05). Comparisons of the POC INRs to the group mean of the POC methods, show higher correlation (R>0.93), but there were still significant (p<0.05) differences noted between the POC group INR mean and CoaguChek Plus, ACT II, TAS PT-One, TAS PTNC, and Hemachron Jr Signature INRs. These data indicate that POC INR biases exist between laboratory methods and POC devices. Until a suitable whole blood INR standardization method is available, we conclude that clinicians using point-of-care anticoagulation monitoring should be aware of differences between POC and parent laboratory values.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Varfarina / Sistemas Automatizados de Assistência Junto ao Leito / Coeficiente Internacional Normatizado / Anticoagulantes Tipo de estudo: Diagnostic_studies Limite: Animals / Humans Idioma: En Revista: Thromb Haemost Ano de publicação: 2000 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Varfarina / Sistemas Automatizados de Assistência Junto ao Leito / Coeficiente Internacional Normatizado / Anticoagulantes Tipo de estudo: Diagnostic_studies Limite: Animals / Humans Idioma: En Revista: Thromb Haemost Ano de publicação: 2000 Tipo de documento: Article País de afiliação: Estados Unidos