Correction of cardiac output obtained by Modelflow from finger pulse pressure profiles with a respiratory method in humans.
Tam, Enrico; Azabji Kenfack, Marcel; Cautero, Michela; Lador, Federic; Antonutto, Guglielmo; di Prampero, Pietro Enrico; Ferretti, Guido; Capelli, Carlo.
Afiliação
  • Tam E; Dipartimento di Scienze e Tecnologie Biomediche, School of Medicine, P.le Kolbe 4, I-33100, Udine, Italy.
Clin Sci (Lond) ; 106(4): 371-6, 2004 Apr.
Article em En | MEDLINE | ID: mdl-14606953
ABSTRACT
The beat-by-beat non-invasive assessment of cardiac output (Q litre x min(-1)) based on the arterial pulse pressure analysis called Modelflow can be a very useful tool for quantifying the cardiovascular adjustments occurring in exercising humans. Q was measured in nine young subjects at rest and during steady-state cycling exercise performed at 50, 100, 150 and 200 W by using Modelflow applied to the Portapres non-invasive pulse wave (Q(Modelflow)) and by means of the open-circuit acetylene uptake (Q(C2H2)). Q values were correlated linearly ( r = 0.784), but Bland-Altman analysis revealed that mean Q(Modelflow) - Q(C2H2) difference (bias) was equal to 1.83 litre x min(-1) with an S.D. (precision) of 4.11 litre x min(-1), and 95% limits of agreement were relatively large, i.e. from -6.23 to +9.89 litre x min(-1). Q(Modelflow) values were then multiplied by individual calibrating factors obtained by dividing Q(C2H2) by Q(Modelflow) for each subject measured at 150 W to obtain corrected Q(Modelflow) (Qcorrected) values. Qcorrected values were compared with the corresponding Q(C2H2) values, with values at 150 W ignored. Data were correlated linearly ( r = 0.931) and were not significantly different. The bias and precision were found to be 0.24 litre x min(-1) and 3.48 litre x min(-1) respectively, and 95% limits of agreement ranged from -6.58 to +7.05 litre x min(-1). In conclusion, after correction by an independent method, Modelflow was found to be a reliable and accurate procedure for measuring Q in humans at rest and exercise, and it can be proposed for routine purposes.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Débito Cardíaco / Monitorização Ambulatorial da Pressão Arterial / Dedos Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Humans / Male Idioma: En Revista: Clin Sci (Lond) Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Itália
Buscar no Google
Base de dados: MEDLINE Assunto principal: Débito Cardíaco / Monitorização Ambulatorial da Pressão Arterial / Dedos Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Humans / Male Idioma: En Revista: Clin Sci (Lond) Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Itália