Analysis of guanidine in high salt and protein matrices by cation-exchange chromatography and UV detection.
J Chromatogr A
; 1073(1-2): 263-7, 2005 May 06.
Article
em En
| MEDLINE
| ID: mdl-15909528
A simple and sensitive HPLC method for quantitative determination of guanidine in high salt and protein matrices was developed. The HPLC system consisted of an Agilent 1100 pump with an online degasser, a UV detector, an autosampler, and Dionex CS 14 cation-exchange guard (4 mm x 50 mm) and analytical (4 mm x 250 mm) columns. The mobile phase was 3.75 mM methanesulfonic acid (MSA) with a flow rate of 1 mL/min. The other analysis parameters were: 50 microL injection volume, 195 nm UV detection, and 21 min runtime. The limit of quantitation (LOQ) for guanidine HCl was determined to be 0.25 mg/L and the standard curve ranged from 0.25 mg/L to 10 mg/L. Sample preparation was required for the samples containing high protein concentrations. Proteins were removed by centrifuging a sample in a 30 K NanoSep centrifugal filter at 15,300 x g for 20 min. The method could determine guanidine accurately in sample matrices containing up to 200 mM sodium ion or up to 50 mM potassium ion. The method can be used for clearance testing of guanidine in biopharmaceutical products.
Buscar no Google
Base de dados:
MEDLINE
Assunto principal:
Sais
/
Espectrofotometria Ultravioleta
/
Cromatografia por Troca Iônica
/
Guanidina
Tipo de estudo:
Diagnostic_studies
Idioma:
En
Revista:
J Chromatogr A
Ano de publicação:
2005
Tipo de documento:
Article
País de afiliação:
Estados Unidos