Your browser doesn't support javascript.
loading
Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths.
Rohrer, Martin; Bauer, Hans; Mintorovitch, Jan; Requardt, Martin; Weinmann, Hanns-Joachim.
Afiliação
  • Rohrer M; Global Business Unit Diagnostic Imaging, Schering AG, Berlin, Germany. Martin.Rohrer@Schering.de
Invest Radiol ; 40(11): 715-24, 2005 Nov.
Article em En | MEDLINE | ID: mdl-16230904
ABSTRACT
RATIONALE AND

OBJECTIVES:

To characterize and compare commercially available contrast media (CM) for magnetic resonance imaging (MRI) in terms of their relaxivity at magnetic field strengths ranging from 0.47 T to 4.7 T at physiological temperatures in water and in plasma. Relaxivities also were quantified in whole blood at 1.5 T.

METHODS:

Relaxivities of MRI-CM were determined by nuclear magnetic resonance (NMR) spectroscopy at 0.47 T and MRI phantom measurements at 1.5 T, 3 T, and 4.7 T, respectively. Both longitudinal (T1) and transverse relaxation times (T2) were measured by appropriate spin-echo sequences. Nuclear magnetic resonance dispersion (NMRD) profiles were also determined for all agents in water and in plasma.

RESULTS:

Significant dependencies of relaxivities on the field strength and solvents were quantified. Protein binding leads to both increased field strength and solvent dependencies and hence to significantly altered T1 relaxivity values at higher magnetic field strengths.

CONCLUSIONS:

Awareness of the field strength and solvent associated with relaxivity data is crucial for the comparison and evaluation of relaxivity values. Data observed at 0.47 T can thus be misleading and should be replaced by relaxivities measured at 1.5 T and at 3 T in plasma at physiological temperature.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Meios de Contraste / Gadolínio / Magnetismo Idioma: En Revista: Invest Radiol Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Alemanha
Buscar no Google
Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Meios de Contraste / Gadolínio / Magnetismo Idioma: En Revista: Invest Radiol Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Alemanha