Your browser doesn't support javascript.
loading
Improving delivery of segments with small MU in step-and-shoot IMRT.
Kuperman, V Y; Lam, W C.
Afiliação
  • Kuperman VY; James A Haley Veterans Hospital, Tampa, FL 33612, USA. vadimkuperman@yahoo.com
Med Phys ; 33(4): 1067-73, 2006 Apr.
Article em En | MEDLINE | ID: mdl-16696483
The purpose of this study is to describe four new delivery schemes for intensity-modulated radiation therapy (IMRT). In the first two schemes the order in which segments are delivered is varied from fraction to fraction. The last two delivery schemes employ fixed order of segments. The obtained results indicate that the suggested approaches can significantly reduce the so-called "overshoot" and "undershoot" phenomena and the associated discrepancies between planned and delivered monitor units.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Radiometria / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Fracionamento da Dose de Radiação / Radioterapia Conformacional / Modelos Biológicos Tipo de estudo: Evaluation_studies Idioma: En Revista: Med Phys Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Radiometria / Algoritmos / Planejamento da Radioterapia Assistida por Computador / Fracionamento da Dose de Radiação / Radioterapia Conformacional / Modelos Biológicos Tipo de estudo: Evaluation_studies Idioma: En Revista: Med Phys Ano de publicação: 2006 Tipo de documento: Article País de afiliação: Estados Unidos